K9840 Auto Kjeldahl Distillation Unit
The K9840 Auto Kjeldahl Distillation Unit is designed to determine the nitrogen content of samples based on the Kjeldahl nitrogen determination method. Fully intelligent software is able to complete sample distillation within minutes, boasts automatic reagent quantification and filling, and automatic testing.
The distillation and condensation automatic cleaning system further enhances measurement precision; explicit interface is easy for operation. It is widely used in food processing, feed production, tobacco, livestock, soil fertility, environmental monitoring, medicine, agriculture, scientific research, teaching, quality control and other fields for the test of nitrogen or protein content, can also be used for the test of ammonium, volatile fatty acid/alkali, and so on.
Key Characteristics
- Display: 4.3-inch color screen
- Operation Modes: Manual/Automatic mode changeover free
- Cleaning System: Automatic Alkali pipeline washing function
- Filling Options: Automatic or manual filling mode according to test need
- Distillation Control: Editable distillation time with automatic alarming upon completion
- Cleaning Automation: Automatic cleaning of control system and distiller
- Safety Design: Intelligent design with safety designs and displacement hints
- Cooling System: Intelligent cooling water control system
- Emergency Features: Emergency stop operation
- Diagnostics: Automatic fault detection and intelligent audible and visual alarm system
- Calibration: Dilution water, Alkali solution, Boric acid solution, and washing flow calibration
- Monitoring: Real-time display of safety door, digestion tube and cooling water conditions
- Self-Testing: Built-in testing model for detecting most functions condition
Advanced Features
- Automatic accurate distribution of alkali and boric acid solution
- Perfect safety protection against over temperature and over pressure
- Comprehensive calibration functions for all solutions
- Real-time monitoring of safety conditions during operation
Industry Applications
Food Processing
Agriculture
Pharmaceuticals
Tobacco
Livestock
Quality Control
Scientific Research
Education
Environmental Monitoring
Soil Fertility
Feed Production
Medical Research
Technical Specifications
Parameter | Specification |
---|---|
Measuring range | 0.1mg ~ 240mg N |
Recovery | ≧99.5% |
Sample capacity | Solid: ≦6g/sample Liquid: ≦16mL/sample |
Distillation speed | 3-6min/sample |
Cooling water consumption | 1.5L/min |
Operating mode | Manual/automatic mode |
Display modes | 4.3″ LCD screen |
Power supply | 220VAC±10%, 50Hz |
Power | 1.3KW |
Net weight | 30Kg |