DPH100/DEC100/DDO100/DPC100/DPD100/DCD100/DPCD100 Multi-Parameter Meter

DPH100 / DEC100 / DDO100 / DPC100 / DPD100 / DCD100 / DPCD100

Benchtop Water Quality Meters — Classical Series with 7″ Display

Professional benchtop water quality meters with high-resolution 7″ backlit display. Supports automatic calibration, ATC measurement, and data export via flash disk or wireless printer. Available in single-parameter and multi-parameter configurations.

Key Features

  • High-res 7″ display screen with white backlight, easy to read
  • Automatic calibration, ATC measurement, parameter setup
  • Automatic reading stable indication, calibration indication
  • Data storage: 450 readings
  • Data export: Flash disk / Wireless printer (the printer is optional)
  • Equipped with a flexible probe holder and pH buffer solutions for convenient use
  • Protection class: IP54

Model Overview

Type Model Parameter Measured
Single Parameter DPH100 pH / mV / Temperature
Single Parameter DEC100 Conductivity / TDS / Salinity / Resistivity / Temperature
Single Parameter DDO100 Dissolved Oxygen / Temperature
Multi-Parameter DPC100 pH / mV / Conductivity / Temperature
Multi-Parameter DPD100 pH / mV / DO / Temperature
Multi-Parameter DCD100 Conductivity / DO / Temperature
Multi-Parameter DPCD100 pH / mV / Conductivity / DO / Temperature

Specifications — Single Parameter

Parameter Specification DPH100 DEC100 DDO100
Parameter pH/mV
mV (ORP)
Cond/TDS/Sal./Res.
DO
pH Measurement Range (-2.00 ~ 20.00) pH
Resolution 0.1 / 0.01
Accuracy ±0.01 pH ±one digit
Calibration Points 1, 2, 3 pts automatic
ATC Range (-9.9 ~ 130.0)°C / (14.18 ~ 266.0)°F
ORP Measurement Range (-1999.9 ~ +1999.9) mV
Resolution 0.1
Accuracy ±0.1% F.S ±one digit
Conductivity Measurement Range 0 ~ 200 mS/cm (Auto-ranging); TDS: 0 ~ 100 g/L; Salinity: 0 ~ 100 ppt; Resistivity: 0 ~ 100 MΩ·cm
Resolution 0.01 / 0.1 / 1
Accuracy ±1.0% F.S ±one digit
DO Measurement Range (0.00 ~ 20.00) mg/L; (0.0 ~ 200.0)%
Resolution 0.01 mg/L; 0.1%
Accuracy ±0.30 mg/L
Temperature Measurement Range (-9.9 ~ 130.0)°C / (14.18 ~ 266.0)°F
Accuracy ±0.2°C

Specifications — Multi-Parameter

Parameter Specification DPC100 DPD100 DCD100 DPCD100
Parameter pH/mV
mV (ORP)
Cond/TDS/Sal./Res.
DO
pH Measurement Range (-2.00 ~ 20.00) pH
Resolution 0.1 / 0.01
Accuracy ±0.01 pH ±one digit
Calibration Points 1, 2, 3 pts automatic
ATC Range (-9.9 ~ 130.0)°C / (14.18 ~ 266.0)°F
ORP Measurement Range (-1999.9 ~ +1999.9) mV
Resolution 0.1
Accuracy ±0.1% F.S ±one digit
Conductivity Measurement Range 0 ~ 200 mS/cm (Auto-ranging); TDS: 0 ~ 100 g/L; Salinity: 0 ~ 100 ppt; Resistivity: 0 ~ 100 MΩ·cm
Resolution 0.01 / 0.1 / 1
Accuracy ±1.0% F.S ±one digit
DO Measurement Range (0.00 ~ 20.00) mg/L; (0.0 ~ 200.0)%
Resolution 0.01 mg/L; 0.1%
Accuracy ±0.30 mg/L
Temperature Measurement Range (-9.9 ~ 130.0)°C / (14.18 ~ 266.0)°F
Accuracy ±0.2°C
Scroll to Top