TopPette Mechanical Pipettes
Features
- Lightweight, ergonomic, low-force design
- Digital display clearly reads volume setting
- Volume range 0.1 μL to 20 mL
- Easy calibration and maintenance with the supplied tool
- Design helps avoid repetitive strain injuries
- Calibrated in accordance with ISO 8655; each pipette has an individual test certificate
- Lower section can be autoclaved
- Turn the plunger button for volume selection
- Comfortable finger support
- Tip ejector for convenient one-handed operation
- Removable ejector collar and tip cone
- Durable tip cone materials provide chemical resistance

Pipette with Switch
DLAB pipettes can be stopped with an on/off switch and cable. This can be used with coagulation analyzers or other instruments requiring precise timing.
TopPette Multi-Channel
- 8- and 12-channel pipettes for standard 96-well plates
- Rotating dispensing head for pipetting convenience
- Individual piston and tip cone assemblies for easy repair and maintenance
- Compound-material tip cone design allows visual seal verification
Pipette Stands
| Stand | Cat. No. | Compatibility and Capacity |
|---|---|---|
| Linear Stand | 7030000085 | Compatible with TopPette, MicroPette Plus and HiPette. Holds up to 8 single-channel or multi-channel pipettes simultaneously, or up to 10 HiPette pipettes |
Single-Channel Adjustable Volume Pipettes
Error limits in accordance with ISO 8655-2.
| Volume Range (μL) | Increment (μL) | Test Volume (μL) | Systematic Error (μL) | Systematic Error (%) | Random Error (μL) | Random Error (%) |
|---|---|---|---|---|---|---|
| 0.1-2.5 | 0.05 | 2.5 | ±0.0625 | ±2.50 | ±0.05 | ±2.00 |
| 0.1-2.5 | 0.05 | 1.25 | ±0.0375 | ±3.00 | ±0.0375 | ±3.00 |
| 0.1-2.5 | 0.05 | 0.25 | ±0.03 | ±12.00 | ±0.015 | ±6.00 |
| 0.5-10 | 0.1 | 10 | ±0.1 | ±1.00 | ±0.08 | ±0.80 |
| 0.5-10 | 0.1 | 5 | ±0.075 | ±1.50 | ±0.075 | ±1.50 |
| 0.5-10 | 0.1 | 1 | ±0.025 | ±2.50 | ±0.015 | ±1.50 |
| 2-20 | 0.5 | 20 | ±0.18 | ±0.90 | ±0.08 | ±0.40 |
| 2-20 | 0.5 | 10 | ±0.12 | ±1.20 | ±0.1 | ±1.00 |
| 2-20 | 0.5 | 2 | ±0.06 | ±3.00 | ±0.04 | ±2.00 |
| 5-50 | 0.5 | 50 | ±0.3 | ±0.60 | ±0.15 | ±0.30 |
| 5-50 | 0.5 | 25 | ±0.225 | ±0.90 | ±0.15 | ±0.60 |
| 5-50 | 0.5 | 5 | ±0.1 | ±2.00 | ±0.1 | ±2.00 |
| 10-100 | 1 | 100 | ±0.8 | ±0.80 | ±0.15 | ±0.15 |
| 10-100 | 1 | 50 | ±0.5 | ±1.00 | ±0.2 | ±0.40 |
| 10-100 | 1 | 10 | ±0.3 | ±3.00 | ±0.15 | ±1.50 |
| 20-200 | 1 | 200 | ±1.2 | ±0.60 | ±0.3 | ±0.15 |
| 20-200 | 1 | 100 | ±0.8 | ±0.80 | ±0.3 | ±0.30 |
| 20-200 | 1 | 20 | ±0.6 | ±3.00 | ±0.2 | ±1.00 |
| 50-200 | 1 | 200 | ±1.2 | ±0.60 | ±0.3 | ±0.15 |
| 50-200 | 1 | 100 | ±0.8 | ±0.80 | ±0.3 | ±0.30 |
| 50-200 | 1 | 50 | ±0.5 | ±1.00 | ±0.2 | ±0.40 |
| 100-1000 | 5 | 1000 | ±6 | ±0.60 | ±2 | ±0.20 |
| 100-1000 | 5 | 500 | ±3.5 | ±0.70 | ±1.25 | ±0.25 |
| 100-1000 | 5 | 100 | ±2 | ±2.00 | ±0.7 | ±0.70 |
| 200-1000 | 5 | 1000 | ±6 | ±0.60 | ±2 | ±0.20 |
| 200-1000 | 5 | 500 | ±3.5 | ±0.70 | ±1.25 | ±0.25 |
| 200-1000 | 5 | 200 | ±1.8 | ±0.90 | ±0.6 | ±0.30 |
| 500-5000 | 50 | 5000 | ±25 | ±0.50 | ±7.5 | ±0.15 |
| 500-5000 | 50 | 2500 | ±15 | ±0.60 | ±7.5 | ±0.30 |
| 500-5000 | 50 | 500 | ±3.5 | ±0.70 | ±1.5 | ±0.30 |
| 1000-5000 | 50 | 5000 | ±25 | ±0.50 | ±7.5 | ±0.15 |
| 1000-5000 | 50 | 2500 | ±15 | ±0.60 | ±7.5 | ±0.30 |
| 1000-5000 | 50 | 1000 | ±7 | ±0.70 | ±3 | ±0.30 |
| 1000-10000 | 100 | 10000 | ±60 | ±0.60 | ±20 | ±0.20 |
| 1000-10000 | 100 | 5000 | ±60 | ±1.20 | ±15 | ±0.30 |
| 1000-10000 | 100 | 1000 | ±30 | ±3.00 | ±6 | ±0.60 |
| 2000-10000 (not applicable to MicroPette Plus) | 100 | 10000 | ±60 | ±0.60 | ±20 | ±0.20 |
| 2000-10000 (not applicable to MicroPette Plus) | 100 | 5000 | ±60 | ±1.20 | ±15 | ±0.30 |
| 2000-10000 (not applicable to MicroPette Plus) | 100 | 2000 | ±60 | ±3.00 | ±12 | ±0.60 |
| *2000-20000 (Coming soon) | 100 | 2000 | ±120 | ±6.00 | ±60 | ±3.00 |
| *2000-20000 (Coming soon) | 100 | 10000 | ±120 | ±1.20 | ±60 | ±0.60 |
| *2000-20000 (Coming soon) | 100 | 20000 | ±120 | ±0.60 | ±60 | ±0.30 |
User calibration should refer to industrial standard ISO 8655-2.
Fixed Volume Pipettes
| Volume Range (μL) | Increment | Test Volume (μL) | Systematic Error (μL) | Systematic Error (%) | Random Error (μL) | Random Error (%) |
|---|---|---|---|---|---|---|
| 5 | – | 5 | ±0.065 | ±1.3 | ±0.06 | ±1.2 |
| 10 | – | 10 | ±0.08 | ±0.8 | ±0.08 | ±0.8 |
| 20 | – | 20 | ±0.12 | ±0.6 | ±0.1 | ±0.5 |
| 25 | – | 25 | ±0.125 | ±0.5 | ±0.075 | ±0.3 |
| 50 | – | 50 | ±0.25 | ±0.5 | ±0.15 | ±0.3 |
| 100 | – | 100 | ±0.5 | ±0.5 | ±0.3 | ±0.3 |
| 200 | – | 200 | ±0.8 | ±0.4 | ±0.4 | ±0.2 |
| 250 | – | 250 | ±1.0 | ±0.4 | ±0.5 | ±0.2 |
| 500 | – | 500 | ±1.5 | ±0.3 | ±1.0 | ±0.2 |
| 1000 | – | 1000 | ±3.0 | ±0.3 | ±2.0 | ±0.2 |
| 2000 | – | 2000 | ±6.0 | ±0.3 | ±3.0 | ±0.15 |
| 5000 | – | 5000 | ±15 | ±0.3 | ±7.5 | ±0.15 |
8-Channel Adjustable Volume Pipettes
| Volume Range (μL) | Increment (μL) | Test Volume (μL) | Systematic Error (μL) | Systematic Error (%) | Random Error (μL) | Random Error (%) |
|---|---|---|---|---|---|---|
| 0.5-10 | 0.1 | 10 | ±0.15 | ±1.50 | ±0.15 | ±1.50 |
| 0.5-10 | 0.1 | 5 | ±0.125 | ±2.50 | ±0.125 | ±2.50 |
| 0.5-10 | 0.1 | 1 | ±0.04 | ±4.00 | ±0.04 | ±4.00 |
| 5-50 | 0.5 | 50 | ±0.5 | ±1.00 | ±0.25 | ±0.50 |
| 5-50 | 0.5 | 25 | ±0.375 | ±1.50 | ±0.25 | ±1.00 |
| 5-50 | 0.5 | 5 | ±0.15 | ±3.00 | ±0.1 | ±2.00 |
| 10-100 | 1 | 100 | ±0.8 | ±0.80 | ±0.3 | ±0.30 |
| 10-100 | 1 | 50 | ±0.5 | ±1.00 | ±0.4 | ±0.80 |
| 10-100 | 1 | 10 | ±0.3 | ±3.00 | ±0.2 | ±2.00 |
| 20-200* | 1 | 200 | ±1.4 | ±0.70 | ±0.5 | ±0.25 |
| 20-200* | 1 | 100 | ±1.0 | ±1.00 | ±0.5 | ±0.50 |
| 20-200* | 1 | 20 | ±0.6 | ±3.00 | ±0.32 | ±1.60 |
| 30-300* | 5 | 300 | ±2.1 | ±0.70 | ±0.75 | ±0.25 |
| 30-300* | 5 | 150 | ±1.5 | ±1.00 | ±0.75 | ±0.5 |
| 30-300* | 5 | 30 | ±0.9 | ±3.00 | ±0.48 | ±1.60 |
| 50-300 | 5 | 300 | ±2.1 | ±0.70 | ±0.75 | ±0.25 |
| 50-300 | 5 | 150 | ±1.5 | ±1.00 | ±0.75 | ±0.50 |
| 50-300 | 5 | 50 | ±0.75 | ±1.50 | ±0.4 | ±0.80 |
| 100-1000 | 10 | 1000 | ±10 | ±1.00 | ±3 | ±0.30 |
| 100-1000 | 10 | 500 | ±10 | ±2.00 | ±2.5 | ±0.50 |
| 100-1000 | 10 | 100 | ±5 | ±5.00 | ±1 | ±1.00 |
12-Channel Adjustable Volume Pipettes
| Volume Range (μL) | Increment (μL) | Test Volume (μL) | Systematic Error (μL) | Systematic Error (%) | Random Error (μL) | Random Error (%) |
|---|---|---|---|---|---|---|
| 0.5-10 | 0.1 | 10 | ±0.15 | ±1.50 | ±0.15 | ±1.50 |
| 0.5-10 | 0.1 | 5 | ±0.125 | ±2.50 | ±0.125 | ±2.50 |
| 0.5-10 | 0.1 | 1 | ±0.04 | ±4.00 | ±0.04 | ±4.00 |
| 5-50 | 0.5 | 50 | ±0.5 | ±1.00 | ±0.25 | ±0.50 |
| 5-50 | 0.5 | 25 | ±0.375 | ±1.50 | ±0.25 | ±1.00 |
| 5-50 | 0.5 | 5 | ±0.15 | ±3.00 | ±0.1 | ±2.00 |
| 10-100 | 0.1 | 200 | ±1.4 | ±0.70 | ±0.5 | ±0.30 |
| 10-100 | 0.1 | 100 | ±1.0 | ±1.00 | ±0.5 | ±0.80 |
| 10-100 | 0.1 | 20 | ±0.6 | ±3.00 | ±0.32 | ±2.00 |
| 20-200* | 1 | 300 | ±2.1 | ±0.70 | ±0.75 | ±0.25 |
| 20-200* | 1 | 150 | ±1.5 | ±1.00 | ±0.75 | ±0.50 |
| 20-200* | 1 | 30 | ±0.9 | ±3.00 | ±0.48 | ±1.60 |
| 30-300* | 5 | 100 | ±0.8 | ±0.80 | ±0.3 | ±0.30 |
| 30-300* | 5 | 50 | ±0.5 | ±1.00 | ±0.4 | ±0.80 |
| 30-300* | 5 | 10 | ±0.3 | ±3.00 | ±0.2 | ±2.00 |
| 50-300 | 5 | 300 | ±2.1 | ±0.70 | ±0.75 | ±0.25 |
| 50-300 | 5 | 150 | ±1.5 | ±1.00 | ±0.75 | ±0.50 |
| 50-300 | 5 | 50 | ±0.75 | ±1.50 | ±0.4 | ±0.80 |
* DLAB specifications are used as guidelines; user calibration should refer to industrial standard ISO 8655.