TopPette Mechanical Pipettes

TopPette Mechanical Pipettes

Features

  • Lightweight, ergonomic, low-force design
  • Digital display clearly reads volume setting
  • Volume range 0.1 μL to 20 mL
  • Easy calibration and maintenance with the supplied tool
  • Design helps avoid repetitive strain injuries
  • Calibrated in accordance with ISO 8655; each pipette has an individual test certificate
  • Lower section can be autoclaved
  • Turn the plunger button for volume selection
  • Comfortable finger support
  • Tip ejector for convenient one-handed operation
  • Removable ejector collar and tip cone
  • Durable tip cone materials provide chemical resistance

TopPette

Pipette with Switch

DLAB pipettes can be stopped with an on/off switch and cable. This can be used with coagulation analyzers or other instruments requiring precise timing.

TopPette Multi-Channel

  • 8- and 12-channel pipettes for standard 96-well plates
  • Rotating dispensing head for pipetting convenience
  • Individual piston and tip cone assemblies for easy repair and maintenance
  • Compound-material tip cone design allows visual seal verification

Pipette Stands

Stand Cat. No. Compatibility and Capacity
Linear Stand 7030000085 Compatible with TopPette, MicroPette Plus and HiPette. Holds up to 8 single-channel or multi-channel pipettes simultaneously, or up to 10 HiPette pipettes

Single-Channel Adjustable Volume Pipettes

Error limits in accordance with ISO 8655-2.

Volume Range (μL) Increment (μL) Test Volume (μL) Systematic Error (μL) Systematic Error (%) Random Error (μL) Random Error (%)
0.1-2.5 0.05 2.5 ±0.0625 ±2.50 ±0.05 ±2.00
0.1-2.5 0.05 1.25 ±0.0375 ±3.00 ±0.0375 ±3.00
0.1-2.5 0.05 0.25 ±0.03 ±12.00 ±0.015 ±6.00
0.5-10 0.1 10 ±0.1 ±1.00 ±0.08 ±0.80
0.5-10 0.1 5 ±0.075 ±1.50 ±0.075 ±1.50
0.5-10 0.1 1 ±0.025 ±2.50 ±0.015 ±1.50
2-20 0.5 20 ±0.18 ±0.90 ±0.08 ±0.40
2-20 0.5 10 ±0.12 ±1.20 ±0.1 ±1.00
2-20 0.5 2 ±0.06 ±3.00 ±0.04 ±2.00
5-50 0.5 50 ±0.3 ±0.60 ±0.15 ±0.30
5-50 0.5 25 ±0.225 ±0.90 ±0.15 ±0.60
5-50 0.5 5 ±0.1 ±2.00 ±0.1 ±2.00
10-100 1 100 ±0.8 ±0.80 ±0.15 ±0.15
10-100 1 50 ±0.5 ±1.00 ±0.2 ±0.40
10-100 1 10 ±0.3 ±3.00 ±0.15 ±1.50
20-200 1 200 ±1.2 ±0.60 ±0.3 ±0.15
20-200 1 100 ±0.8 ±0.80 ±0.3 ±0.30
20-200 1 20 ±0.6 ±3.00 ±0.2 ±1.00
50-200 1 200 ±1.2 ±0.60 ±0.3 ±0.15
50-200 1 100 ±0.8 ±0.80 ±0.3 ±0.30
50-200 1 50 ±0.5 ±1.00 ±0.2 ±0.40
100-1000 5 1000 ±6 ±0.60 ±2 ±0.20
100-1000 5 500 ±3.5 ±0.70 ±1.25 ±0.25
100-1000 5 100 ±2 ±2.00 ±0.7 ±0.70
200-1000 5 1000 ±6 ±0.60 ±2 ±0.20
200-1000 5 500 ±3.5 ±0.70 ±1.25 ±0.25
200-1000 5 200 ±1.8 ±0.90 ±0.6 ±0.30
500-5000 50 5000 ±25 ±0.50 ±7.5 ±0.15
500-5000 50 2500 ±15 ±0.60 ±7.5 ±0.30
500-5000 50 500 ±3.5 ±0.70 ±1.5 ±0.30
1000-5000 50 5000 ±25 ±0.50 ±7.5 ±0.15
1000-5000 50 2500 ±15 ±0.60 ±7.5 ±0.30
1000-5000 50 1000 ±7 ±0.70 ±3 ±0.30
1000-10000 100 10000 ±60 ±0.60 ±20 ±0.20
1000-10000 100 5000 ±60 ±1.20 ±15 ±0.30
1000-10000 100 1000 ±30 ±3.00 ±6 ±0.60
2000-10000 (not applicable to MicroPette Plus) 100 10000 ±60 ±0.60 ±20 ±0.20
2000-10000 (not applicable to MicroPette Plus) 100 5000 ±60 ±1.20 ±15 ±0.30
2000-10000 (not applicable to MicroPette Plus) 100 2000 ±60 ±3.00 ±12 ±0.60
*2000-20000 (Coming soon) 100 2000 ±120 ±6.00 ±60 ±3.00
*2000-20000 (Coming soon) 100 10000 ±120 ±1.20 ±60 ±0.60
*2000-20000 (Coming soon) 100 20000 ±120 ±0.60 ±60 ±0.30

User calibration should refer to industrial standard ISO 8655-2.

Fixed Volume Pipettes

Volume Range (μL) Increment Test Volume (μL) Systematic Error (μL) Systematic Error (%) Random Error (μL) Random Error (%)
5 – 5 ±0.065 ±1.3 ±0.06 ±1.2
10 – 10 ±0.08 ±0.8 ±0.08 ±0.8
20 – 20 ±0.12 ±0.6 ±0.1 ±0.5
25 – 25 ±0.125 ±0.5 ±0.075 ±0.3
50 – 50 ±0.25 ±0.5 ±0.15 ±0.3
100 – 100 ±0.5 ±0.5 ±0.3 ±0.3
200 – 200 ±0.8 ±0.4 ±0.4 ±0.2
250 – 250 ±1.0 ±0.4 ±0.5 ±0.2
500 – 500 ±1.5 ±0.3 ±1.0 ±0.2
1000 – 1000 ±3.0 ±0.3 ±2.0 ±0.2
2000 – 2000 ±6.0 ±0.3 ±3.0 ±0.15
5000 – 5000 ±15 ±0.3 ±7.5 ±0.15

8-Channel Adjustable Volume Pipettes

Volume Range (μL) Increment (μL) Test Volume (μL) Systematic Error (μL) Systematic Error (%) Random Error (μL) Random Error (%)
0.5-10 0.1 10 ±0.15 ±1.50 ±0.15 ±1.50
0.5-10 0.1 5 ±0.125 ±2.50 ±0.125 ±2.50
0.5-10 0.1 1 ±0.04 ±4.00 ±0.04 ±4.00
5-50 0.5 50 ±0.5 ±1.00 ±0.25 ±0.50
5-50 0.5 25 ±0.375 ±1.50 ±0.25 ±1.00
5-50 0.5 5 ±0.15 ±3.00 ±0.1 ±2.00
10-100 1 100 ±0.8 ±0.80 ±0.3 ±0.30
10-100 1 50 ±0.5 ±1.00 ±0.4 ±0.80
10-100 1 10 ±0.3 ±3.00 ±0.2 ±2.00
20-200* 1 200 ±1.4 ±0.70 ±0.5 ±0.25
20-200* 1 100 ±1.0 ±1.00 ±0.5 ±0.50
20-200* 1 20 ±0.6 ±3.00 ±0.32 ±1.60
30-300* 5 300 ±2.1 ±0.70 ±0.75 ±0.25
30-300* 5 150 ±1.5 ±1.00 ±0.75 ±0.5
30-300* 5 30 ±0.9 ±3.00 ±0.48 ±1.60
50-300 5 300 ±2.1 ±0.70 ±0.75 ±0.25
50-300 5 150 ±1.5 ±1.00 ±0.75 ±0.50
50-300 5 50 ±0.75 ±1.50 ±0.4 ±0.80
100-1000 10 1000 ±10 ±1.00 ±3 ±0.30
100-1000 10 500 ±10 ±2.00 ±2.5 ±0.50
100-1000 10 100 ±5 ±5.00 ±1 ±1.00

12-Channel Adjustable Volume Pipettes

Volume Range (μL) Increment (μL) Test Volume (μL) Systematic Error (μL) Systematic Error (%) Random Error (μL) Random Error (%)
0.5-10 0.1 10 ±0.15 ±1.50 ±0.15 ±1.50
0.5-10 0.1 5 ±0.125 ±2.50 ±0.125 ±2.50
0.5-10 0.1 1 ±0.04 ±4.00 ±0.04 ±4.00
5-50 0.5 50 ±0.5 ±1.00 ±0.25 ±0.50
5-50 0.5 25 ±0.375 ±1.50 ±0.25 ±1.00
5-50 0.5 5 ±0.15 ±3.00 ±0.1 ±2.00
10-100 0.1 200 ±1.4 ±0.70 ±0.5 ±0.30
10-100 0.1 100 ±1.0 ±1.00 ±0.5 ±0.80
10-100 0.1 20 ±0.6 ±3.00 ±0.32 ±2.00
20-200* 1 300 ±2.1 ±0.70 ±0.75 ±0.25
20-200* 1 150 ±1.5 ±1.00 ±0.75 ±0.50
20-200* 1 30 ±0.9 ±3.00 ±0.48 ±1.60
30-300* 5 100 ±0.8 ±0.80 ±0.3 ±0.30
30-300* 5 50 ±0.5 ±1.00 ±0.4 ±0.80
30-300* 5 10 ±0.3 ±3.00 ±0.2 ±2.00
50-300 5 300 ±2.1 ±0.70 ±0.75 ±0.25
50-300 5 150 ±1.5 ±1.00 ±0.75 ±0.50
50-300 5 50 ±0.75 ±1.50 ±0.4 ±0.80

* DLAB specifications are used as guidelines; user calibration should refer to industrial standard ISO 8655.

Scroll to Top