TopPette Mechanical Pipettes
TopPette mechanical pipettes are designed for accurate and ergonomic liquid handling across a wide volume range.
Features
- Simply turn the plunger button for volume selection
- Light weight, ergonomic, low force design
- Comfortable finger support
- Digital display clearly reads volume setting
- Tip ejector allows convenient one-handed operation
- The pipettes cover volume range of 0.1μL to 20mL
- Easy to calibrate and maintain with tool supplied
- Design helps avoid repetitive strain injuries
- Calibrated in accordance with ISO8655. Each pipette supplied with individual test certificate
- The lower section can be autoclaved
- Durable tip cone materials provide excellent chemical resistance
- Ejector collar and tip cone can be removed
- Pipette with on/off switch and cable. This can be used with coagulation analyzers or any other instruments that require precise timing
TopPette Multi-channel
TopPette multi-channel pipettes are ideal for work with 96-well plates.
Features
- 8 and 12 channel pipettes are available for standard 96-well plates
- Dispensing head rotates for optimum pipetting convenience
- Individual piston and tip cone assemblies allow easy repair and maintenance
- Compound material tip cone design allows visual seal verification
Pipette Stands
Linear Stand — Cat. No. 7030000085
Compatible with TopPette, MicroPette Plus, and HiPette pipette series. The stand can stably and conveniently hold up to 8 single-channel or multi-channel pipettes simultaneously, and up to 10 pipettes for the HiPette series.
Compatible with TopPette, MicroPette Plus, and HiPette pipette series. The stand can stably and conveniently hold up to 8 single-channel or multi-channel pipettes simultaneously, and up to 10 pipettes for the HiPette series.
Round Stand — Cat. No. 7030000084
Compatible with MicroPette Plus pipette series. The freely rotatable circular stand can securely hold up to 6 single-channel or multi-channel pipettes.
Compatible with MicroPette Plus pipette series. The freely rotatable circular stand can securely hold up to 6 single-channel or multi-channel pipettes.
Specifications
Error limits in accordance with ISO 8655-2.
Single-channel Adjustable Volume Pipettes
| Volume Range μL |
Increment μL |
Test Volume μL |
Systematic Error | Random Error | ||
|---|---|---|---|---|---|---|
| μL | % | μL | % | |||
| 0.1-2.5 | 0.05 | 2.5 | ±0.0625 | ±2.50 | ±0.05 | ±2.00 |
| 1.25 | ±0.0375 | ±3.00 | ±0.0375 | ±3.00 | ||
| 0.25 | ±0.03 | ±12.00 | ±0.015 | ±6.00 | ||
| 0.5-10 | 0.1 | 10 | ±0.1 | ±1.00 | ±0.08 | ±0.80 |
| 5 | ±0.075 | ±1.50 | ±0.075 | ±1.50 | ||
| 1 | ±0.025 | ±2.50 | ±0.015 | ±1.50 | ||
| 2-20 | 0.5 | 20 | ±0.18 | ±0.90 | ±0.08 | ±0.40 |
| 10 | ±0.12 | ±1.20 | ±0.1 | ±1.00 | ||
| 2 | ±0.06 | ±3.00 | ±0.04 | ±2.00 | ||
| 5-50 | 0.5 | 50 | ±0.3 | ±0.60 | ±0.15 | ±0.30 |
| 25 | ±0.225 | ±0.90 | ±0.15 | ±0.60 | ||
| 5 | ±0.1 | ±2.00 | ±0.1 | ±2.00 | ||
| 10-100 | 1 | 100 | ±0.8 | ±0.80 | ±0.15 | ±0.15 |
| 50 | ±0.5 | ±1.00 | ±0.2 | ±0.40 | ||
| 10 | ±0.3 | ±3.00 | ±0.15 | ±1.50 | ||
| 20-200 | 1 | 200 | ±1.2 | ±0.60 | ±0.3 | ±0.15 |
| 100 | ±0.8 | ±0.80 | ±0.3 | ±0.30 | ||
| 20 | ±0.6 | ±3.00 | ±0.2 | ±1.00 | ||
| 50-200 | 1 | 200 | ±1.2 | ±0.60 | ±0.3 | ±0.15 |
| 100 | ±0.8 | ±0.80 | ±0.3 | ±0.30 | ||
| 50 | ±0.5 | ±1.00 | ±0.2 | ±0.40 | ||
| 100-1000 | 5 | 1000 | ±6 | ±0.60 | ±2 | ±0.20 |
| 500 | ±3.5 | ±0.70 | ±1.25 | ±0.25 | ||
| 100 | ±2 | ±2.00 | ±0.7 | ±0.70 | ||
| 200-1000 | 5 | 1000 | ±6 | ±0.60 | ±2 | ±0.20 |
| 500 | ±3.5 | ±0.70 | ±1.25 | ±0.25 | ||
| 200 | ±1.8 | ±0.90 | ±0.6 | ±0.30 | ||
| 500-5000 | 50 | 5000 | ±25 | ±0.50 | ±7.5 | ±0.15 |
| 2500 | ±15 | ±0.60 | ±7.5 | ±0.30 | ||
| 500 | ±3.5 | ±0.70 | ±1.5 | ±0.30 | ||
| 1000-5000 | 50 | 5000 | ±25 | ±0.50 | ±7.5 | ±0.15 |
| 2500 | ±15 | ±0.60 | ±7.5 | ±0.30 | ||
| 1000 | ±7 | ±0.70 | ±3 | ±0.30 | ||
| 1000-10000 | 100 | 10000 | ±60 | ±0.60 | ±20 | ±0.20 |
| 5000 | ±60 | ±1.20 | ±15 | ±0.30 | ||
| 1000 | ±30 | ±3.00 | ±6 | ±0.60 | ||
| 2000-10000* | 100 | 10000 | ±60 | ±0.60 | ±20 | ±0.20 |
| 5000 | ±60 | ±1.20 | ±15 | ±0.30 | ||
| 2000 | ±60 | ±3.00 | ±12 | ±0.60 | ||
| 2000-20000** | 100 | 2000 | ±120 | ±6.00 | ±60 | ±3.00 |
| 10000 | ±120 | ±1.20 | ±60 | ±0.60 | ||
| 20000 | ±120 | ±0.60 | ±60 | ±0.30 | ||
* 2000-10000 μL does not apply to MicroPette plus.
** 2000-20000 μL is coming soon.
User calibration should refer to the industrial standard ISO 8655.
Fixed Volume Pipettes
| Volume Range μL |
Increment | Test Volume μL |
Systematic Error | Random Error | ||
|---|---|---|---|---|---|---|
| μL | % | μL | % | |||
| 5 | – | 5 | ±0.065 | ±1.3 | ±0.06 | ±1.2 |
| 10 | – | 10 | ±0.08 | ±0.8 | ±0.08 | ±0.8 |
| 20 | – | 20 | ±0.12 | ±0.6 | ±0.1 | ±0.5 |
| 25 | – | 25 | ±0.125 | ±0.5 | ±0.075 | ±0.3 |
| 50 | – | 50 | ±0.25 | ±0.5 | ±0.15 | ±0.3 |
| 100 | – | 100 | ±0.5 | ±0.5 | ±0.3 | ±0.3 |
| 200 | – | 200 | ±0.8 | ±0.4 | ±0.4 | ±0.2 |
| 250 | – | 250 | ±1.0 | ±0.4 | ±0.5 | ±0.2 |
| 500 | – | 500 | ±1.5 | ±0.3 | ±1.0 | ±0.2 |
| 1000 | – | 1000 | ±3.0 | ±0.3 | ±2.0 | ±0.2 |
| 2000 | – | 2000 | ±6.0 | ±0.3 | ±3.0 | ±0.15 |
| 5000 | – | 5000 | ±15 | ±0.3 | ±7.5 | ±0.15 |
8-channel Adjustable Volume Pipettes
| Volume Range μL |
Increment μL |
Test Volume μL |
Systematic Error | Random Error | ||
|---|---|---|---|---|---|---|
| μL | % | μL | % | |||
| 0.5-10 | 0.1 | 10 | ±0.15 | ±1.50 | ±0.15 | ±1.50 |
| 5 | ±0.125 | ±2.50 | ±0.125 | ±2.50 | ||
| 1 | ±0.04 | ±4.00 | ±0.04 | ±4.00 | ||
| 5-50 | 0.5 | 50 | ±0.5 | ±1.00 | ±0.25 | ±0.50 |
| 25 | ±0.375 | ±1.50 | ±0.25 | ±1.00 | ||
| 5 | ±0.15 | ±3.00 | ±0.1 | ±2.00 | ||
| 10-100 | 1 | 100 | ±0.8 | ±0.80 | ±0.3 | ±0.30 |
| 50 | ±0.5 | ±1.00 | ±0.4 | ±0.80 | ||
| 10 | ±0.3 | ±3.00 | ±0.2 | ±2.00 | ||
| 20-200* | 1 | 200 | ±1.4 | ±0.70 | ±0.5 | ±0.25 |
| 100 | ±1.0 | ±1.00 | ±0.5 | ±0.50 | ||
| 20 | ±0.6 | ±3.00 | ±0.32 | ±1.60 | ||
| 30-300* | 5 | 300 | ±2.1 | ±0.70 | ±0.75 | ±0.25 |
| 150 | ±1.5 | ±1.00 | ±0.75 | ±0.50 | ||
| 30 | ±0.9 | ±3.00 | ±0.48 | ±1.60 | ||
| 50-300 | 5 | 300 | ±2.1 | ±0.70 | ±0.75 | ±0.25 |
| 150 | ±1.5 | ±1.00 | ±0.75 | ±0.50 | ||
| 50 | ±0.75 | ±1.50 | ±0.4 | ±0.80 | ||
| 100-1000 | 10 | 1000 | ±10 | ±1.00 | ±3 | ±0.30 |
| 500 | ±10 | ±2.00 | ±2.5 | ±0.50 | ||
| 100 | ±5 | ±5.00 | ±1 | ±1.00 | ||
12-channel Adjustable Volume Pipettes
| Volume Range μL |
Increment μL |
Test Volume μL |
Systematic Error | Random Error | ||
|---|---|---|---|---|---|---|
| μL | % | μL | % | |||
| 0.5-10 | 0.1 | 10 | ±0.15 | ±1.50 | ±0.15 | ±1.50 |
| 5 | ±0.125 | ±2.50 | ±0.125 | ±2.50 | ||
| 1 | ±0.04 | ±4.00 | ±0.04 | ±4.00 | ||
| 5-50 | 0.5 | 50 | ±0.5 | ±1.00 | ±0.25 | ±0.50 |
| 25 | ±0.375 | ±1.50 | ±0.25 | ±1.00 | ||
| 5 | ±0.15 | ±3.00 | ±0.1 | ±2.00 | ||
| 10-100 | 1 | 200 | ±1.0 | ±0.70 | ±0.5 | ±0.30 |
| 100 | ±0.5 | ±1.00 | ±0.5 | ±0.80 | ||
| 20 | ±0.6 | ±3.00 | ±0.32 | ±2.00 | ||
| 20-200* | 1 | 200 | ±2.1 | ±0.70 | ±0.75 | ±0.25 |
| 150 | ±1.5 | ±1.00 | ±0.75 | ±0.50 | ||
| 30 | ±0.9 | ±3.00 | ±0.48 | ±1.60 | ||
| 30-300* | 5 | 100 | ±0.8 | ±0.80 | ±0.3 | ±0.30 |
| 50 | ±0.5 | ±1.00 | ±0.4 | ±0.80 | ||
| 10 | ±0.3 | ±3.00 | ±0.2 | ±2.00 | ||
| 50-300 | 5 | 300 | ±2.1 | ±0.70 | ±0.75 | ±0.25 |
| 150 | ±1.5 | ±1.00 | ±0.75 | ±0.50 | ||
| 50 | ±0.75 | ±1.50 | ±0.4 | ±0.80 | ||