400F/300F Series Benchtop Multi-Parameter Meter

GENERAL FEATURES

● 300F Series Benchtop Metesr: High resolution LCD display, 5.7 inches.
● 400F Series Benchtop Metesr: Color high resolution LCD display, 5.7 inches.
● Multi-reading feature allows auto-read, timed-read and continuous-read.
● Automatic/Manual temperature compensation ensures accurate results.
● Auto-hold feature senses and locks the measurement endpoint.
● Data storage 500 sets (GLP-compliant).
● Support for USB or RS-232 communication.
● Reset feature automatically resumes all settings back to factory default options.
● IP54 waterproof.

pH

● 1-5 points calibration with standard recognition.
● PH400F: Selectable pH buffer groups, including NIST, DIN, GB,
USA, MERK.
● PH300F: Selectable pH buffer groups, including NIST, DIN, GB.
● Automatic electrode diagnosis with pH slope and offset display.

Conductivity

● 1-3 points calibration with standard recognition.
● Settable parameters, including cell constant, temperature
compensation coefficient and TDS factor.
● Temperature compensation type (none, linear, pure water).

Ion

● 1-5 points calibration.
● Selectable measurement unit, including μg/L, mg/L, g/L, mmol/L, pX, etc.
● I400F: Multi-measurement modes are supported, including Direct
Reading mode, Standard Addition mode, Sample Addition mode and GRAN mode.
● I300F: Measurement mode is Direct Reading mode.
● Over 10 methods are built-in, including F-, Cl-, Br-, I-, NO3-, BF4-,NH4+, K+, Na+, Ca2+, Cu2+, Pb2+, Ag+ and etc.; user-defined method is supported.

DO

● Support for air-saturated water or zero oxygen calibration.
● Auto barometric pressure compensation.
● Manual Salinity Factor Correction.
● Selectable pressure unit, including kPa, mbar, Torr, Atm.

Model M300F M300F-A PH300F EC300F I300F PH400F EC400F I400F DO400F
Parameters pH/EC/ISE/DO/Temp. (mV/pX/Resistivity/TDS/Sal/ DO Saturation) pH/EC/ISE/Temp. (mV/pX/Resistivity/ TDS/Sal.) pH/Temp. (mV) EC/Temp. (Resistivity/ TDS/Sal.) pH/ISE/Temp. (mV/pX) pH/Temp. (mV) EC/Temp. (Resistivity/TDS/ Sal.) pH/ISE/Temp. (mV/pX) DO/Temp. (DO Saturation)
pH Range -2.00 to 20.00pH / -2.000 to 20.000 pH / /
Resolution 0.1, 0.01 pH / 0.1, 0.01, 0.001pH / /
Accuracy ±0.01 pH / ±0.002 pH / /
Calibration Points Up to 5 / / /
Standard Customization Yes / / /
Standard Recognition NIST, GB and DIN buffers / NIST,GB,DIN,USA
and MERK buffers
/ /
mV Range -2000.0 to 2000.0 mV / -2000.00 to 2000.00 mV / /
Resolution 0.1 mV / 0.1, 0.01 mV / /
Accuracy ±0.3 mV or ±0.1 % of reading whichever is greater / ±0.1 mV or ±0.03 %
of reading whichever is greater
/ /
pX Range -2.00 to 20.00 / / / / -2.000 to 20.000 pX /
Resolution 0.1, 0.01 pX / / / / 0.1, 0.01, 0.001 pX /
Accuracy ±0.02 pX / / / / ±0.002 pX /
Calibration Points Up to 5 / / / / /
ISE Range 1E-9 to 9.999E9 / / / / /
Unit mol/L, mmol/L, g/L, mg/L, μg/L, ppm / / / / /
Resolution Up to 4 significant digits / / / / /
Accuracy ±0.5 % / / / / ±0.3 % /
Calibration Points Up to 5 / / / / /
Conductivity Range 0.000 μS/cm to 1000 mS/cm / / / / /
Resolution 0.001 μS/cm minimum; changed with range / / / / /
Accuracy ±1.0 % FS / / / ±0.5 % FS / /
Reference Temperature 20, 25 ºC / / / / /
Calibration Points Up to 3 / / / / /
Standard Recognition 84μS/cm, 1413μS/cm; 12.88mS/cm / / / / /
Resistivity Range 5.00Ω·cm ~ 20.00MΩ·cm / / / 5.00Ω·cm~100.00MΩ·cm / /
Resolution 0.01Ω·cm  minimum / / / / /
Accuracy ±1.0 % FS / / / ±0.5 % FS / /
TDS Range 0.00mg/L ~ 300g/L / / / 0.00mg/L~1000g/L / /
Resolution 0.01mg/L minimum; changed with range / / / 0.001mg/L minimum; changed with range / /
Accuracy ±1.0% FS / / / ±0.5 % FS / /
Salinity Range (0.00 ~ 8.00)% / / / / /
Resolution 0.01% / / / / /
Accuracy ±0.2% / / / ±0.1% / /
Dissolved Oxygen Concentration Sensor Type Polarographic / / / / / / /
Range 0.00 to 20.00 mg/L / / / / / / / 0.00 to 99.99mg/L
Resolution 0.01mg/L / / / / / / /
Accuracy ±0.3mg/L / / / / / / / ±0.3mg/L (0.00 to 20.00 mg/L)
±10% (20.00 to 99.99 mg/L)
Calibration Points Air-saturated water or zero point / / / / / / /
Barometric Compensation yes / / / / / / /
Manual Salinity Factor Correction yes / / / / / / /
Saturation Range (0.0 to 200.0)% / / / / / / / (0.0 to 600.0) %
Resolution 0.1% / / / / / / /
Accuracy ±10.0 % / / / / / / /
Temperature Range -5 to 110 ºC, 23 to 230 ºF -10 to 135 ºC,
14 to 275 ºF
Unit ºC,  ºF ºC,  ºF
Resolution 0.1 0.1
Accuracy ±0.2 ±0.1
Measurement Reading Mode AutoRead( Fast, Medium, Slow), Timed, Continuous
Reading Prompts Reading, Stable, Locked
Temp. Compensation ATC, MTC
Data Management Data Storage 500 results each
GLP Features Yes
Inputs pH Electrode BNC(Q9) / / /
Temp./DO Probe 4-pin aviation connector / /
Temp. /EC Probe 5-pin aviation connector / / / / /
Outputs USB PC
RS 232 printer
Display Options Backlight Yes
Auto Shut-down 1~60 min, off
IP Rating IP54
Date and Time Yes
General Power AC Adapter,100-240 VAC input, DC9V output
Dimensions 242 x 195 x  68 mm
Weight 900g(1.98 lb)
SKU: 400f-300f-series Category: Tags: ,

Catalog

Download Catalog