- 53l meter: Conductivity, Salinity, Temperature.
Multi-parameter portable conductivity meter, with TDS, salinity, resistivity measurement mode, 5-point button calibration, setting menu can customize 11 functional parameters, measurement accuracy: 0.5% FS.
Features:
• Multi-parameter portable conductivity meter, equipped with a backlit LCD display.
• 1 to 5 point calibration, automatic identification of conductivity standards.
• Optional cell constant (K=0.1/1/10), reference temperature (20/25°C), TDS conversion coefficient, linearity and pure water compensation, seawater and practical
salinity measurement modes.
• Automatic temperature compensation, correcting the conductivity measurement and converting the reading to the reference temperature.
• Automatic end point lock, keep stable readings for easy browsing and recording.
• Calibration due reminder, reminding the user to calibrate the meter regularly.
• Calibration report, showing calibration points and conductivity cell coefficients.
• Automatic power off, effectively saving battery power consumption.
• The setting menu can customize the number of calibration points, stability conditions, temperature units, date and time, etc.
• Reset function, automatically restore the meter to the factory default settings.
• Expand memory, store or recall up to 500 sets of data.
• USB communication interface, easy to transfer data to a personal computer.
• Multi-mode power supply scheme (battery, power adapter, computer USB port) to ensure the smooth application of the meter.
Specifications
model | Bante531 |
Conductivity | |
Measuring range | 0.01~20.00, 200.0, 2000µS/cm, 20.00, 200.0mS/cm |
display resolution | 0.001, 0.01, 0.1, 1 |
measurement accuracy | ±0.5% FS |
Calibration point | 1 to 5 points |
Calibration solution | 10µS, 84µS, 1413µS, 12.88mS, 111.8mS |
salinity | |
Measuring range | 0.00~80.00ppt, 0.00~42.00psu |
display resolution | 0.01 |
measurement accuracy | ±1% FS |
temperature | |
Measuring range | 0~105°C/32~221°F |
display resolution | 0.1°C/0.1°F |
measurement accuracy | ±0.5°C/±0.9°F |
Offset calibration | 1 point, measured value ±10°C |
General parameters | |
Temperature compensation range | 0~100°C/32~212°F, manual or automatic |
Temperature compensation coefficient | Linear (0.0~10.0%/°C), pure water |
Reference temperature | 20/25°C |
Conductivity cell constant | K=0.1, 1, 10 or customized |
Stability conditions | Standard or high precision |
Calibration due reminder | 1 to 31 days or closed |
Automatic shut-down | 10, 20, 30 minutes after no key operation |
Data storage | 500 groups |
Communication output | USB |
Connector | 6-pin DIN, 3.5mm microphone jack |
Display screen | Custom LCD display (80×60mm) |
Power type | 3 AA batteries or DC5V power adapter |
Dimensions | 170(L)×85(W)×30(H)mm |
Meter weight | 300g |