400F/300F Series Benchtop Multi-Parameter Meter

GENERAL FEATURES

● 300F Series Benchtop Metesr: High resolution LCD display, 5.7 inches.
● 400F Series Benchtop Metesr: Color high resolution LCD display, 5.7 inches.
● Multi-reading feature allows auto-read, timed-read and continuous-read.
● Automatic/Manual temperature compensation ensures accurate results.
● Auto-hold feature senses and locks the measurement endpoint.
● Data storage 500 sets (GLP-compliant).
● Support for USB or RS-232 communication.
● Reset feature automatically resumes all settings back to factory default options.
● IP54 waterproof.

pH

● 1-5 points calibration with standard recognition.
● PH400F: Selectable pH buffer groups, including NIST, DIN, GB,
USA, MERK.
● PH300F: Selectable pH buffer groups, including NIST, DIN, GB.
● Automatic electrode diagnosis with pH slope and offset display.

Conductivity

● 1-3 points calibration with standard recognition.
● Settable parameters, including cell constant, temperature
compensation coefficient and TDS factor.
● Temperature compensation type (none, linear, pure water).

Ion

● 1-5 points calibration.
● Selectable measurement unit, including μg/L, mg/L, g/L, mmol/L, pX, etc.
● I400F: Multi-measurement modes are supported, including Direct
Reading mode, Standard Addition mode, Sample Addition mode and GRAN mode.
● I300F: Measurement mode is Direct Reading mode.
● Over 10 methods are built-in, including F-, Cl-, Br-, I-, NO3-, BF4-,NH4+, K+, Na+, Ca2+, Cu2+, Pb2+, Ag+ and etc.; user-defined method is supported.

DO

● Support for air-saturated water or zero oxygen calibration.
● Auto barometric pressure compensation.
● Manual Salinity Factor Correction.
● Selectable pressure unit, including kPa, mbar, Torr, Atm.

Model M300F M300F-A PH300F EC300F I300F PH400F EC400F I400F DO400F
Parameters pH/EC/ISE/DO/Temp. (mV/pX/Resistivity/TDS/Sal/ DO Saturation) pH/EC/ISE/Temp. (mV/pX/Resistivity/ TDS/Sal.) pH/Temp. (mV) EC/Temp. (Resistivity/ TDS/Sal.) pH/ISE/Temp. (mV/pX) pH/Temp. (mV) EC/Temp. (Resistivity/TDS/ Sal.) pH/ISE/Temp. (mV/pX) DO/Temp. (DO Saturation)
pH Range -2.00 to 20.00pH ● ● / ● -2.000 to 20.000 pH / ● /
Resolution 0.1, 0.01 pH ● ● / ● 0.1, 0.01, 0.001pH / ● /
Accuracy ±0.01 pH ● ● / ● ±0.002 pH / ● /
Calibration Points Up to 5 ● ● / ● ● / ● /
Standard Customization Yes ● ● / ● ● / ● /
Standard Recognition NIST, GB and DIN buffers ● ● / ● NIST,GB,DIN,USA
and MERK buffers
/ ● /
mV Range -2000.0 to 2000.0 mV ● ● / ● -2000.00 to 2000.00 mV / ● /
Resolution 0.1 mV ● ● / ● 0.1, 0.01 mV / ● /
Accuracy ±0.3 mV or ±0.1 % of reading whichever is greater ● ● / ● ±0.1 mV or ±0.03 %
of reading whichever is greater
/ ● /
pX Range -2.00 to 20.00 ● / / ● / / -2.000 to 20.000 pX /
Resolution 0.1, 0.01 pX ● / / ● / / 0.1, 0.01, 0.001 pX /
Accuracy ±0.02 pX ● / / ● / / ±0.002 pX /
Calibration Points Up to 5 ● / / ● / / ● /
ISE Range 1E-9 to 9.999E9 ● / / ● / / ● /
Unit mol/L, mmol/L, g/L, mg/L, μg/L, ppm ● / / ● / / ● /
Resolution Up to 4 significant digits ● / / ● / / ● /
Accuracy ±0.5 % ● / / ● / / ±0.3 % /
Calibration Points Up to 5 ● / / ● / / ● /
Conductivity Range 0.000 μS/cm to 1000 mS/cm ● / ● / / ● / /
Resolution 0.001 μS/cm minimum; changed with range ● / ● / / ● / /
Accuracy ±1.0 % FS ● / ● / / ±0.5 % FS / /
Reference Temperature 20, 25 ºC ● / ● / / ● / /
Calibration Points Up to 3 ● / ● / / ● / /
Standard Recognition 84μS/cm, 1413μS/cm; 12.88mS/cm ● / ● / / ● / /
Resistivity Range 5.00Ω·cm ~ 20.00MΩ·cm ● / ● / / 5.00Ω·cm~100.00MΩ·cm / /
Resolution 0.01Ω·cm  minimum ● / ● / / ● / /
Accuracy ±1.0 % FS ● / ● / / ±0.5 % FS / /
TDS Range 0.00mg/L ~ 300g/L ● / ● / / 0.00mg/L~1000g/L / /
Resolution 0.01mg/L minimum; changed with range ● / ● / / 0.001mg/L minimum; changed with range / /
Accuracy ±1.0% FS ● / ● / / ±0.5 % FS / /
Salinity Range (0.00 ~ 8.00)% ● / ● / / ● / /
Resolution 0.01% ● / ● / / ● / /
Accuracy ±0.2% ● / ● / / ±0.1% / /
Dissolved Oxygen Concentration Sensor Type Polarographic / / / / / / / ●
Range 0.00 to 20.00 mg/L / / / / / / / 0.00 to 99.99mg/L
Resolution 0.01mg/L / / / / / / / ●
Accuracy ±0.3mg/L / / / / / / / ±0.3mg/L (0.00 to 20.00 mg/L)
±10% (20.00 to 99.99 mg/L)
Calibration Points Air-saturated water or zero point / / / / / / / ●
Barometric Compensation yes / / / / / / / ●
Manual Salinity Factor Correction yes / / / / / / / ●
Saturation Range (0.0 to 200.0)% / / / / / / / (0.0 to 600.0) %
Resolution 0.1% / / / / / / / ●
Accuracy ±10.0 % / / / / / / / ●
Temperature Range -5 to 110 ºC, 23 to 230 ºF ● ● ● ● -10 to 135 ºC,
14 to 275 ºF
● ● ●
Unit ºC,  ºF ● ● ● ● ºC,  ºF ● ● ●
Resolution 0.1 ● ● ● ● 0.1 ● ● ●
Accuracy ±0.2 ● ● ● ● ±0.1 ● ● ●
Measurement Reading Mode AutoRead( Fast, Medium, Slow), Timed, Continuous ● ● ● ● ● ● ● ●
Reading Prompts Reading, Stable, Locked ● ● ● ● ● ● ● ●
Temp. Compensation ATC, MTC ● ● ● ● ● ● ● ●
Data Management Data Storage 500 results each ● ● ● ● ● ● ● ●
GLP Features Yes ● ● ● ● ● ● ● ●
Inputs pH Electrode BNC(Q9) ● ● / ● ● / ● /
Temp./DO Probe 4-pin aviation connector ● ● / ● ● / ● ●
Temp. /EC Probe 5-pin aviation connector ● / ● / / ● / /
Outputs USB PC ● ● ● ● ● ● ● ●
RS 232 printer ● ● ● ● ● ● ● ●
Display Options Backlight Yes ● ● ● ● ● ● ● ●
Auto Shut-down 1~60 min, off ● ● ● ● ● ● ● ●
IP Rating IP54 ● ● ● ● ● ● ● ●
Date and Time Yes ● ● ● ● ● ● ● ●
General Power AC Adapter,100-240 VAC input, DC9V output ● ● ● ● ● ● ● ●
Dimensions 242 x 195 x  68 mm ● ● ● ● ● ● ● ●
Weight 900g(1.98 lb) ● ● ● ● ● ● ● ●
Scroll to Top