The high-precision portable pH ion meter contains 3 measurement functions. The meter is suitable for measuring the pH, oxidation-reduction potential and ion concentration of the sample. The measurement accuracy is 0.002pH.
Features of pH mode:
• High-precision portable pH ion meter with backlit LCD display.
• 1 to 5 point calibration, automatic identification of USA/NIST/DIN standard buffers.
• Automatic electrode diagnosis, display pH slope and zero offset.
ORP mode features:
• 1-point offset calibration, allowing adjustment of the displayed value to a known standard.
• Relative and absolute mV modes to ensure reliable redox potential measurement.
Features of ion concentration mode:
• 2 to 5 point calibration, including 8 concentration calibration points for selection.
• Directly read the ion concentration value to avoid tedious conversion and calculation.
• Automatic electrode diagnosis, with calibration point and slope display.
• Optional concentration units (ppm, mg/L, mol/L) and ion media.
Common features:
• Automatic temperature compensation to ensure accurate measurement of the full range.
• Automatic end point lock, keep stable readings for easy browsing and recording.
• Calibration due reminder, reminding the user to calibrate the meter regularly.
• Automatic power off, effectively saving battery power consumption.
• The setting menu can customize the number of calibration points, display resolution, stability conditions, temperature units, date and time, etc.
• Reset function, automatically restore the meter to the factory default settings.
• Expand memory, store or recall up to 500 sets of data.
• USB communication interface, easy to transfer data to a personal computer.
• Multi-mode power supply scheme (battery, power adapter, computer USB port) to ensure the smooth application of the meter.
model | Bante320 |
pH | |
Measuring range | -2.000~20.000pH |
display resolution | 0.001, 0.01, 0.1pH, optional |
measurement accuracy | ± 0.002pH |
Calibration point | 1 to 5 points |
pH buffer option | USA, NIST, DIN, 2 custom buffers |
ORP | |
Measuring range | ±1999.9mV |
display resolution | 0.1, 1mV, optional |
measurement accuracy | ±0.2mV |
Calibration point | 1 o’clock |
Measurement mode | Relative or absolute mV |
Ion concentration | |
Measuring range | 0.001~19999 (depending on the range of ion selective electrode) |
display resolution | 0.001, 0.01, 0.1, 1 |
measurement accuracy | ±0.5% FS (one intermediary ion), ±1% FS (two intermediary ion) |
Calibration point | 2 to 5 o’clock |
Calibration solution | 0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000 |
Units of measurement | ppm, mg/L, mol/L, mmol/L |
temperature | |
Measuring range | 0~105°C/32~221°F |
display resolution | 0.1°C/0.1°F |
measurement accuracy | ±0.5°C/±0.9°F |
Offset calibration | 1 point, measured value ±10°C |
General parameters | |
Temperature compensation range | 0~100°C/32~212°F, manual or automatic |
Data lock | Manual or automatic end point lock |
Stability conditions | Standard or high precision |
Calibration due reminder | 1 to 31 days or closed |
Automatic shut-down | 10, 20, 30 minutes after no key operation |
Data storage | 500 groups |
Communication output | USB |
Connector | BNC, 3.5mm microphone jack |
Display screen | Custom LCD display (80×60mm) |
Power type | 3 AA batteries or DC5V power adapter |
Dimensions | 170(L)×85(W)×30(H)mm |
Meter weight | 300g |