CO2 Incubator (Infrared Sensor) — Standard Range
The new BPN series CO2 incubators are upgraded replacements for the original HH·CP series. Combining nearly twenty years of research, development and manufacturing experience in this field, they are guided by user needs and continuously incorporate newly researched technologies. Infrared CO2 sensors provide accurate, stable control unaffected by temperature and humidity.
Product Features
User-Friendly Design
- Units can be stacked in two tiers to make full use of laboratory space. A microcomputer LCD controller provides precise, reliable control. The stainless-steel chamber and shelves have rounded corners and freely removable shelf supports for easy chamber cleaning;
Infrared Sensor
The infrared (IR) sensor rapidly monitors CO2 gas and is unaffected by fluctuations in external temperature and humidity. Even with multiple users and frequent door opening and closing, chamber CO2 concentration remains stable and uniform. Thermal-conductivity sensors, in contrast, are susceptible to temperature and humidity, have long response times and large detection errors, and require frequent calibration during long-term use;
Microbial Filter
- The CO2 inlet is fitted with a high-efficiency microbial filter. For particles with diameters greater than or equal to 0.3μm, its filtration efficiency is up to 99.99%, effectively filtering bacteria and dust particles from the CO2 gas;
Automatic Circulating Fan Speed Control
- Circulating fan speed is automatically controlled. When the chamber reaches a constant temperature, the speed decreases to an airflow suitable for cell growth, preventing sample evaporation caused by excessive airflow during experiments;
Door Heating System
- The incubator door heats the inner glass door to effectively prevent condensation and the potential microbial contamination caused by it;
Safety Functions
- An independent temperature-limit alarm system (optional) ensures safe experiments without accidents;
- Low-temperature, high-temperature, overtemperature and prolonged door-opening alarms;
UV Sterilization System (UV Series)
- The UV sterilization lamp at the rear top of the chamber periodically disinfects the interior, effectively killing airborne bacteria in the circulating air and water vapor from the humidification tray, thereby preventing contamination during cell culture;
Technical Specifications
| Technical Specification / Model | BPN-50CH(UV) BPN-80CH(UV) |
BPN-150CH(UV) BPN-150CHH |
BPN-190CH(UV) BPN-190CHH |
BPN-240CH(UV) BPN-240CHH |
BPN-30CW(UV) BPN-80CW(UV) |
BPN-150CW(UV) |
|---|---|---|---|---|---|---|
| Supply Voltage | AC220V 50HZ | |||||
| Input Power | 450W/500W | 750W | 750W | 950W | 250W/680W | 950W |
| Heating Method | Air-jacketed | Water-jacketed | ||||
| Temperature Control Range | RT+5~55°C | |||||
| Ambient Operating Temperature | +5~30°C | |||||
| Temperature Fluctuation | ±0.2°C | ±0.1°C | ||||
| CO2Control Range | 0-20% | |||||
| CO2 Control Precision | ±0.1% (infrared sensor) | |||||
| CO2 Recovery Time | (After door opening for 30 seconds, recovery to 5%) ≤3 minutes | |||||
| Temperature Recovery | (After door opening for 30 seconds, recovery to 37°C) ≤8 minutes | |||||
| Relative Humidity | Natural evaporation ≥90% | |||||
| Capacity | 50L/80L | 150L | 190L | 240L | 26L/80L | 150L |
| Chamber Dimensions (mm) W×D×H |
400×350×350 400×450×500 |
480×530×610 | 520×530×690 | 600×630×670 | 290×290×310 400×400×500 |
500×500×650 |
| Overall Dimensions (mm) W×D×H |
580×450×730 590×657×870 |
670×710×950 | 708×710×1030 | 788×837×940 | 440×410×544 550×520×764 |
650×615×914 |
| Shelves (Standard) | 2 shelves | 3 shelves | 3 shelves | 3 shelves | 2 shelves | 3 shelves |
| Disinfection Method | UV disinfection / 90°C high-temperature, high-humidity sterilization (CHH series) | UV disinfection | ||||
90°C high-temperature moist-heat sterilization system (CHH series): sterilizes the chamber, including the temperature sensor, CO2 concentration sensor, fan, shelves and supports, with high-temperature moist heat. This eliminates microbial contamination of cell cultures caused by bacteria, mold, mycoplasma and other microorganisms, providing a safe experimental environment.
Optional Accessories:
1. RS485 interface and communication software
2. Dedicated CO2 pressure-reducing regulator
3. Humidity display system
4. Independent temperature-limit controller
5. Built-in printer
6. HEPA filter
Note 1: choose either RS485/232 or printer.
Note 2: 3Q qualification and calibration services are available.
Note 3: All technical data are measured at an ambient temperature of 25°C;