the ssr series digital rotary viscometer adopts coaxial cylinder structure, using small sample adapter, the sample quantity required for each measurement is very small, only 7ml-13ml (rotor 0 is 21ml), optional constant temperature water jacket with cycle, with constant temperature tank to control the sample under test. The operation is very convenient and the measurement is accurate. very suitable for precious, rare sample measurement.
characteristic:
1. Timing measurement function
2. Coaxial cylinder structure
3. Using a spindle rotor, the sample size required for testing is very small
4. Two RS232 interfaces with direct connection to micro printer and computer
1.Detailed technical parameters of LV-SSR rotary viscometer:
model | LV-SSR |
display mode | digital display |
speed(r/min) | 0.3/0.6/1.5/3/6/10/12/15/20/25/30/40/50/60/70/80/90/100
Total 18 |
rotor | 18#,25#,31#,34# (standard configuration)
ULR (0#) (apolegamy) |
measuring range | 1- 1,6M (mPa﹒s) |
Sample size | ULR(0# rotor): 1-1000; 21ml
18# rotor:3 –10K; 7ml 25# rotor:480 –1.6M; 9ml 31# rotor:30 –100K; 10.5ml 34# rotor:60 –200K; 11ml |
measurement error | ±3% (Newtonian liquid) |
repetitive error | ±1.5% (Newtonian liquid) |
output interface | Two RS232 interfaces to connect printer and computer |
Temperature Probe Interface | standard configuration |
outline dimension | 300 × 300 × 450(mm) |
- Detailed technical parameters of the RV-SSR / HA-SSR / HB-SSR rotary viscometer:
model | RV-SSR | HA-SSR | HB-SSR |
display mode | digital display | ||
speed(r/min) | 0.5/1/2/2.5/4/5/10/15/20/25/30/40/50/60/70/80/90/100
Total 18 |
||
rotor | 21#、27#、28#、29# (standard configuration)
ULR (0#) (apolegamy) |
||
measuring range | 6.4 – 2M
0# rotor:6.4-1K 21# rotor:50-100K 27# rotor:250-500K 28# rotor:500-1M 29# rotor:1K-2M |
12.8 – 4M
0# roto:12.8-1K 21# roto:100-200K 27# roto:500-1M 28# roto:1K-2M 29# roto:2K-4M |
51.2 – 16M
0# roto:51.2-2K 21# roto子:400-800K 27# roto:2K-4M 28# roto:4K-8M 29# roto:8K-16M |
Sample size | ULR(0# roto):21ml
21# roto: 7.8ml 27# roto: 11.3ml 28# roto: 12.6ml 29# roto: 11.5ml |
||
measurement error | ±3% (Newtonian liquid) | ||
repetitive error | ±1.5% (Newtonian liquid) | ||
output interface | Two RS232 interfaces to connect printer and computer | ||
Temperature Probe Interface | standard configuration | ||
outline dimension | 300 × 300 × 450(mm) |
Accessories:
Ultra Low Viscosity Adapter (ULR)
The Ultra Low Viscosity Adapter Rotor (ULR for short) is used with our Viscometer to provide accurate and repeatable measurements on low viscosity objects. Both Newtonian and non-Newtonian fluids are suitable. The ULR consists of a precision cylindrical rotor and a precision tubular sleeve (sample cartridge). The rheology-adapted cylindrical design allows for precise viscosity measurement and shear rate measurement.
- Designed for low viscosity fluid measurement
- Available in interlayer and non-interlayer versions
- The minimum detection limit is 1cP, which depends on the type of viscometer used.
- Accurate measurement at specified shear rate for detailed analysis of products
- Stainless steel material is easy to clean and clean, can withstand the operating temperature range of -40℃-200℃
- Sample dosage 21 mL