LS-60FS Series Vertical Pressure Steam Sterilizers
Manual Flip-top | Technical Data


Characteristics
- Flip-top design, rod type manual open and close the cover, more efficient and prevent the infection.
- (Optional) Equipped with a drying function, which is provided by a drying silica gel belt to enhance the drying effect of sterilized items.
- (Optional) An external vacuum pump can be connected. Once connected with the vacuum pump, it can become a sterilizer with pulsating vacuum function, and users can switch freely.
- 7" LCD touch screen with Android operating system, enabling convenient operation of the device.
- The whole process of water inlet, (vacuum), heat, sterilization, drainage, steam discharge, and (drying) is proceed automatically.
- Reservation function: The date, year, month, day, hour and minute can be set.
- The operation interface is equipped with multiple fixed programs and melting/holding programs. It also features a dedicated test program for pharmaceutical packaging materials. Additionally, there are three user-defined programs available, allowing users to select the appropriate program based on the items to be sterilized.
- 0.22 \mu m high-efficiency air filter ensures the incoming air is sterile and avoids recontamination.
- Steam-water inner circulation system: No steam discharge out, it ensures a clean and dry environment.
- Standard PT/TT test port is convenient to carry out the future testing.
- Chamber material: 2.5mm SUS304 stainless steel made, with double-layer stainless steel mesh baskets.
- Temperature and pressure: Maximum working temperature: 138°C, Design pressure 0.3Mpa, Design temperature: 150°C.
- Time range: Sterilization time: 1–9999 minutes, Melting time: 1–9999 minutes, Preservation time: 1–9999 minutes.
- USB storage function: the USB port is provided, so the sterilizing data can be stored conveniently to be viewed and traced.
- GMP-compliant authority management and data storage: with more than three levels of password authority control, each level has an independent password, it can memorize and store 100 sterilization cycles and support expandable storage.
- Safety protection devices: Electric and mechanical dual safety interlocks, lid closure safety inspection system, automatic fault detection system, over-temperature protection, low-water protection,
Technical Data
| MODEL Technical Data Chamber volume |
LS-60FS | LS-80FS | LS-100FS | LS-120FS |
|---|---|---|---|---|
| 60L Φ 396 × 490mm | 80L Φ 396 × 650mm | 100L Φ 480 × 560mm | 120L Φ 480 × 660mm | |
| The height after rising | 1310mm | 1590mm | 1475mm | 1565mm |
| Design pressure | 0.3MPa | |||
| Design temperature | 150°C | |||
| Max. working pressure | 0.25MPa | |||
| Max. working temperature | 138°C | |||
| Heat average | ≤ ± 1°C | |||
| Timer range | 0~99min | |||
| Power | (3+1.3)KW/AC220V.50Hz | (4.5+1.3)KW/AC220V.50Hz | (5.8+1.3)KW/AC220V.50Hz | |
| Overall dimension(mm) | 686×556×913 | 686×556×1035 | 840×600×970 | 840×600×1060 |
| Transport dimension(mm) | 786×656×1013 | 786×656×1135 | 940×610×1000 | 940×610×1075 |
| G.W/N.W | 210kg/180kg | 230kg/200kg | 300kg/270kg | 325kg/295kg |