LHH-150GSD-UV Series Combined Pharmaceutical Photostability Test Chamber

Category: Sort:

catalog_LHH-150GSD-UV_LHH-150GSP-UV_LHH-250GSD-UV_LHH-250GSP-UV_LHH-500GSD-UV_44_SHYH

Application Overview

Yiheng’s combined pharmaceutical photostability test chamber (with UV monitoring and control) adds visible-light and ultraviolet monitoring and control to the original GSD/GSP/GP combined stability test chambers. It is essential equipment for GMP certification in pharmaceutical and cosmetics companies and meets the photostability testing conditions of the ICH guidelines.
It scientifically establishes long-term stable temperature, humidity and lighting conditions for drug shelf-life evaluation. Suitable for accelerated, long-term, high-humidity and intense-light exposure tests of medicines and new drugs in pharmaceutical companies, it provides a reliable solution for pharmaceutical stability testing.

Product Features

Lighting System of the Combined Pharmaceutical Photostability Test Chamber

  • The lighting system meets the lCH Q1B photostability testing requirements for new drug substances and products and corresponding international manufacturing standards, as well as the 2020 Pharmacopoeia guidelines for pharmaceutical stability testing. It scientifically simulates the long-term stable temperature, humidity and photostability environment needed to evaluate drug shelf life. It meets pharmaceutical companies’ temperature, humidity and photostability testing requirements for new drug substances and products;
  • Shelf-mounted or outer-door lighting can be selected, including visible-light and UV lamps controlled independently. Single- or double-tier visible-light or UV lamps can be selected, and the sample shelf height in the chamber is adjustable (double-tier downward lighting system optional).

Irradiance Display, Monitoring and Control

  • Overcomes the inability of existing pharmaceutical stability chambers to display and monitor irradiance, reducing stability-test errors caused by irradiance decay as visible-light and UV lamps age. Light intensity can also be adjusted continuously to meet test requirements. Visible-light and UV measurement probes with optical sensors, and an irradiance monitor certified by a third party, are provided for convenient observation and calibration;

Specialized UV Lamps

  • Specialized UV lamps meet the ICH Q1B photostability testing requirements for new drug substances and products. Compared with other UV lamps, they provide stable quality and uniform spectral power. Their spectral power distribution does not decline as the lamps age, allowing more repeatable test results;
  • UV lamp holders suitable for high-humidity operation;

User-Friendly Design

  • New CFC-free design: high efficiency and low energy consumption promote energy savings and keep you at the forefront of healthy living;
  • A microcomputer controller provides stable, accurate, reliable control. The 304 stainless-steel chamber has rounded corners for easy cleaning and convenient operation.
  • Air-duct circulation ensures uniform airflow distribution in the chamber;
  • A 25mm-diameter test port is standard on the left side;

Continuous Operation

  • Two imported compressors switch automatically to ensure prolonged continuous pharmaceutical testing without faults, overcoming the limitation of domestically produced pharmaceutical test chambers that cannot operate continuously for extended periods;
  • Continuous operation requires no defrosting, avoiding chamber temperature and humidity fluctuations caused by defrosting during use;

Quality Assurance

  • Key components, including the temperature/humidity controller, compressors and circulating fan, are imported for stable, safe, reliable continuous operation over extended periods;

Imported Humidity Sensor

  • A humidity sensor suitable for high-temperature operation avoids frequent replacement of wet-bulb wicks;

Programmable Touchscreen Controller (Standard on the “GSP” Series)

  • The large touchscreen is easy to operate and program;
  • Chinese and English controller interfaces are selectable, and real-time operating curves can be displayed on the screen;
  • Capacity for 100 program groups, 1000 segments and 999 cycle steps, with a maximum time setting of 99 hours 59 minutes per segment;
  • After data and test conditions are entered, screen locking prevents shutdown caused by accidental touch;
  • P.I.D automatic calculation immediately corrects temperature and humidity change conditions for more precise, stable control;
  • An RS-232 or RS-485 communication interface allows programs to be designed on a computer, testing to be monitored, and start-up, shutdown and other functions to be performed;

Recording and Fault-diagnosis Display

  • If the test chamber develops a fault, the dynamic display shows clear operating fault information;
  • A printer or 485 communication interface can be connected to record temperature and time curves using a computer and printer, supporting storage and playback of experimental data;

Wireless Alarm System (SMS Alarm System) — Optional

  • If the equipment user is away when a fault occurs, the system promptly collects fault signals and sends an SMS to the designated recipient’s mobile phone. This supports prompt fault correction and resumption of testing, preventing losses;

Safety Functions

  • An independent temperature-limit alarm system provides audible and visual warnings, ensuring safe experiments without accidents;
  • Low-temperature, high-temperature and overtemperature alarms, plus high- and low-humidity alarms;
  • A lockable door is standard to prevent accidental opening during tests and UV injury to laboratory personnel;

Technical Specifications

Technical Specification / Model LHH-150GSD-UV
LHH-150GSP-UV
LHH-250GSD-UV
LHH-250GSP-UV
LHH-500GSD-UV
LHH-500GSP-UV
LHH-800GSD-UV
LHH-800GSP-UV
LHH-1000GSD-UV
LHH-1000GSP-UV
LHH-1500GSP-UV
LHH-150GP-UV
LHH-250GP-UV
LHH-400GP-UV
Temperature Control Range 0~65°C (15~50°C with lighting)
Temperature Fluctuation / Uniformity ±0.5°C /±2°C~±3°C
Humidity Range / Deviation 25~95%RH/±3%RH Not provided
Light Intensity / Error Cool-white fluorescent light: 0-6000Lux adjustable / ±500Lux; UV light intensity adjustable
UV Spectral Range 320~400nm
Applicable Test Standard ICH Q1B directive; cool-white fluorescent light ≥1.2×106Lux.hr; near-UV fluorescent light ≥200 w.hr/m2
Light-source Configuration Cool-white fluorescent lamps (meeting ISO 18909) and UVA near-UV fluorescent lamps
Standard Light Sources Cool-white lamps (D65, 6500K) and UV lamps with a wavelength range of 320-400nm
Refrigeration System / Method Two independent original imported hermetic compressors with automatic changeover
Controller Programmable Touchscreen Controller
Power Supply AC220V 50HZ AC220V 50HZ AC380V 50HZ AC220V 50HZ
Operating Power Approx. 680W Approx. 750W Approx. 1400W Approx. 2880W Approx. 2960W / 4800W Approx. 450W / 450W / 600W
Rated Power 2100W 2500W 3910W 3600W 3700W/6000W 1450W/1700W/3200W
Capacity 150L 250L 500L 800L 1000L/1500L 150L/250L/400L
Chamber Dimensions (mm) W×D×H 550×405×670 600×500×830 670×725×1020 800×590×1650 1050×590×1650
1550×590×1650
550×405×670
600×500×830
700×550×1140
Overall Dimensions (mm) W×D×H 690×805×1530 740×890×1680 850×1100×1930 1360×890×2000 1610×890×2000
2110×890×2000
690×805×1530
740×890×1680
950×850×1850
Lighting Method Top lighting / door lighting Top lighting Door lighting / top lighting
Built-in printer Standard
Safety Devices Compressor overheating protection, fan overheating protection, overtemperature protection, compressor overpressure protection, overload protection, water-shortage protection, wired monitoring alarm system (optional)
Remarks 1. High-precision digital recorder optional
2. GP and GSD: manual stepless dimming, standard illuminance monitor and UV irradiance monitor, built-in top UV lighting or door lighting
3. GSP-UV: automatic UV monitoring system with optical-sensor visible-light and UV measurement probes
4. GSD-UV can be fitted with optional multi-tier lighting.

 

Scroll to Top