L3S/L3 Split Beam VIS Spectrophotometer

  • Split beam optical system
  • 8-inch multi color touch screen
  • Automatic zero and full scale adjustment
  • Thermal plotter optional
  • Automatic wavelength settings
  • automatic 8 position cuvette holder
  • Equipped with USB port
  • Software optional

Common Features
1.  Legend series spectrometers adopt Split beam survey optical system, and blazed holographic gratings. They have outstanding test precision and very competitive prices.
2. 8-inch multi color touch-screen, cutting-edge user interface, powerful functions, and easy operation.
3. The automated 8 position cuvette holder facilitates the operation under each measurement mode.
4. With powerful functions, the equipment shows great performance in qualitative and quantitative testing, such as:
● Full-spectrum scanning
● Linear regression
● Detailed spectrum scanning
● Concentration direct reading
● Time-based kinetics determination
● Peak / Valley detecting
● GOTO λ
● Multi-wavelength measurement
5. The equipment is designed with sophisticated power protection system. With high capacity of internal memory, it can store testing results, scanned images, regression equations and correction data. Therefore, it follows a fast initialization when power on.
6. The instrument can be connected with dedicated printer, which can print testing results, or draw curve from spectral scanning, fixed wavelength time-based scanning, and linear regression.
7. With USB COM port the device can be connected to a PC, which can not only enhance the performance in data testing and spectrum scanning, but also expand the memory to save more testing results.

Specifications

L6S L6 L3S L3
Photometry  Split Beam
Monochromator Type Czerny-Turner
Focal Length 160mm
Grating 1200 lines/mm
Detector Silicon Photocell
Spectrum Bandwidth 2nm
Wavelength Setting 8-inch color touch-screen
Wavelength Range 190-1100nm 325-1100nm
Wavelength Accuracy ±0.5nm
Wavelength Repeatability ≤0.2nm
Photometric Range 0.0-200.0%T
-0.301-4.000A
0.000-9999C
Photometric Accuracy ±0.3%T
±0.002Abs (0-0.5A)
±0.004 Abs (0.5-1A)
±0.5%T
Photometric Repeatability ≤0.15%T
0.001 Abs (0-0.5A)
0.002 Abs (0.5-1A)
≤0.2%T
Scanning Speed Fast-Medium-Slow Fast-Medium-Slow
Baseline ≤±0.002A
(200
1090nm)
≤±0.002A
(335-1090nm)
Stray Light ≤0.03%T
( at 220nm NaI,
360nm NaNo2 )
≤0.05%T
( at 220nm NaI,
360nm NaNo2 )
≤0.05%T
( at 360nm NaNo2 )
Noise 100%(T)noise≤0.15%(T) ,
0%(T)noise≤0.1%(T)
100%(T)noise≤0.2%(T) ,
0%(T)noise≤0.1%(T)
Drifting ±0.0009 Abs/30min
(at 250nm and 500nm after 2h warm up)
±0.0009 Abs/30min
(at 500nm after 2h warm up)
Power AC220V±22V 50Hz±1Hz,200W AC220V±22V 50Hz±1Hz,120W
Packaging Size 710mm X 590mm X 505mm 0.21M 3 40kg

 

Standard Accessories
● Operation manual 1
● Glass cuvette 1cm 4pcs
● Quartz cuvette 1cm 2pcs
● Power cable 1
● Fuse 2pcs
● Cuvette holder 10mm

Optional Accessories
● UV WIN8 Spectrum data processing software
● Thermal Plotter
● Cuvette holder 100mm
● Cuvette holder 50m

Catalog