GR60DA Vertical Autoclave

To meet the requirements of a future lab
GR series intelligent autoclaves-a once-off selection that exempts the worries of continual upgrading

Main Features:

Heavy-gauge sterilization chamber

The heavy gauge sterilization chamber in a diameter of 40cm offers a massive space to satisfy various sterilization requirements.

Microcomputer control system

The latest “inspiration” fast-speed microcomputer intelligent control system is highly robust to not only realize the total-process controls of sterilization, but also enable convenient maintenance, inquiry, calibration, recording and upward scalability in the future. It allows three tiers of administrations, to guarantee satisfactory us and management of the autoclave.

Waste sterilization mode

Dedicated waste sterilization procedures are established for effective sterilization of lab wastes.

Drying

Thoroughly drying according to the CE standard, on hot sale in the Europe

Cooling lock OPEN temperature

Each procedure can set different cooling lock OPEN temperatures according to the thermal inertia of the sterilization object so as to prevent the user from opening the cover and hence getting burnt when the sterilization object(particularly liquid) is still in an unsafe state.

Sterilizing temperature

105-138℃

Simple and easily understandable and operable control panel

Preset Procedures and memory system

Six sterilization procedures and warming procedures are preset for the sterilizing, warming and drying of solids and liquids, the melting and warming of agars. There are up to 60 cycles and be created and saved to be readily used.

Safety

Mechanical interlocking: the unique “eight-column evenly distributed” interlocking device is safer and more reliable than common interlocking devices. The electric double inner locks enable the control system to monitor the temperature and pressure in the sterilization chamber on a real-time basis and disable opening of the sterilization chamber cover until the safe temperature and pressure are reached, hence guarantee the personal safety of the operator.

SpecGR60DRGR60DFGR60DA
GR85DRGR85DFGR85DA
GR110DRGR110DFGR110DA
Sterilization modeSolid mode:
Heating-Sterilizing-Exhausting and air cooling
YESYESYES
Solid withdrying mode:
Heating-Sterilizing-Exhaustingand air cooling-water drainage-drying
YESNONO
Liquid mode:
Heating-sterilizing-Precooling-Steam exhausting
YESYESYES
Waste Mode:
Heating-Sterilizing-Steam exhausting and air cooling
YESYESYES
Waste Mode:Heating-Sterilizing-Steam exhausting and air coolingYESYESYES
Agar Mode:
Heating-Melting-Warming
YESYESYES
Automatic water feedingSTANDARDSTANDARDNO
Water level sensorSTANDARDSTANDARDNO
Cooling fanSTANDARDSTANDARDOPTIONAL
Load thermometerOPTIONALOPTIONALOPTIONAL
PrinterOPTIONALOPTIONALOPTIONAL
Memory systemCreate 60 proceduresCreate 60 proceduresCreate 20 procedures

Specification Table

ModeGR60DAGR85DAGR110DA
GR60DFGR85DFGR110DF
GR60DRGR85DRGR110DR
Model differenceDA:Standard, DF:Automatic water feeding+cooling fan, DR: DF+drying
Capacity(L)6085110
Dimension(L*W*H,mm)660x644x980(DA)660x644x980(DA)660x644x1180(DA)
660x831x980(DF/DR)660x831x980(DF/DR)660x831x1180(DF/DR)
Chamber dimension(Dia*H,mm)φ400×505φ400×700φ400×895
Rate power supply(w)290046004600
Stainless steel basket(Dia*H,mm)(φ360×200)x2(φ360×280)x2(φ360×250)x3
Power requirement220V,16A,50/60HZ220V,32A,50/60HZ220V,32A,50/60HZ
Temperature accuracy0.1℃
Working environment5-45℃,relatively humidity 10%-85%
Chamber materialSUS 304
Sterilization temperature105-138℃
Sterilization time range1-6000 minutes
Melting temperature60-100℃
Melting time range1-6000 minutes
Warming temperature45-60℃
Warming time range1-9999 minutes
Drying(only for DR model)1-300 minutes, drying temperature 80-140℃
Cooling lock OPEN temperature40-80℃ for liquid cycle, 40-99℃ for other cycles
Sterilization cycle1.Solid cycle 2.Liquid cycle 3.Liquid with warming cycle 4.Agar cycle 5.Solid with drying(for DR) 6. Medical waste cycle 7. Customized cycle
Sterilization cycle numberDA can save up to 20 cycles, DF/DR can save up to 60 cycles
Exhaust modeAutomatic internal discharging ,with a steam collecting bottle inside. Exhaust speed can be selected by 6 levels
Lid opening wayUp and down
Steam collector and water tankDA: embedded 2 steam collector, DF: embedded 2 steam collector and 28L water tank  DR: embedded steam collector and 28L water tank
Preset range of timerCan be preset and auto start up to 6 days in advance
Display range of pressure gauge-0.1-0.5Mpa
PressureMax working pressure:0.27Mpa,design pressure:0.35Mpa
Controller“Inspiration” fast speed microcomputer controller
Safety device“eight column evenly distributed” interlocking device, electric double inner locks, lid closing checking, dry scorch protection, overpressure protection, safety valve, water level sensor, over temperature protection, temperature monitor, over current and short circuit protection, leakage protection, cooling lock, anti-scald chamber cover and bench, automatic troubleshooting