CRY-3B Syncretize Electric suitable for cell hybridization and cell fusion and direct observations can be made under inverted microscopes. It has a widespread application in the research on microorganism, animal medicine and bio-engineering. It provides safety, ease of use, and efficiency as compared with the conventional method. All types of even electrode and needle electrode are
FEATURES
- Features: both cell electrofusion and electroporation capabilities
- Flexibility: have a wide range of voltage and pulse time
- Fast and efficient, simple and quickly complete post-processing
- process cell arrangement, fusion, fusion, only a few seconds
- LCD touch display operation
- Data setting easy and quickly
TECHNICAL PARAMETERS
AC Electric Field Parameters | ||||||||
Bunchy pulse voltage | Peak-peak value 0-58V(0-±29V), adjustable continuously | |||||||
Bunchy pulse frequency |
30KHz-3000KHz adjustable | |||||||
Gate time | 10-1000 um, 1 us increasing decline, adjustable continuously |
|||||||
DC Field Parameters | ||||||||
Square wave fusion pulse voltage |
5-600V adjustable(electrode distance)1.0mm electric voltage field reach 6000V/cm. | |||||||
Square wave fusion pulse range |
5-5000 – us, 1 us increasing decline, adjustable continuously | |||||||
Fusion (square wave) pulse number |
1-9 pcs | |||||||
With fusion flat electrode |
Polar distance 0.5mm, optional electrode in the table below |
|||||||
Net weight | 9.22Kg | |||||||
Packing size | 450×370×290mm | |||||||
Model | 1-0.5 | 1-1.0 | 1-2.0 | 1-3.2 | ||||
Electrode distance |
0.5mm | 1.0mm | 2mm | 3.2mm | ||||
Volume | 20µL | 40µL | 80µL | 700µL | ||||
Voltage range | 0-600V DC |
APPLICATION AREA
- Cell fusion (the fusion process can be observed under inverted microscope)
- Nuclear transfer
- Embryo operation
- Hybridoma generation
- Plant protoplast fusion
- The chicken embryo genes into inside an egg
- Live gene/drug import
Optional Microscope Slide Electrode
Model | 1-0.5 | 1-1.0 | 1-2.0 | 1-3.2 | ||||
Electrode distance |
0.5mm | 1.0mm | 2mm | 3.2mm | ||||
Volume | 20µL | 40µL | 80µL | 700µL | ||||
Voltage range | 0-600V DC | |||||||
Maintenance | Gentle cleanser, ethanol or ethylene oxide sterilization |