LHH-500SD Series Large Combined Pharmaceutical Stability Test Chamber

Category: Sort:

catalog_LHH-500SD_LHH-500SDP_LHH-800SD_LHH-800SDP_LHH-1000SD_LHH-1000SDP_43-1_SHYH

Scientifically establishes long-term stable temperature, humidity and lighting conditions for drug shelf-life evaluation. Suitable for accelerated, long-term, high-humidity and intense-light exposure tests of medicines and new drugs in pharmaceutical companies, it provides a reliable solution for pharmaceutical stability testing.

Product Features

  • New CFC-free design offers high efficiency and low energy consumption to promote energy savings;
  • An antibacterial stainless-steel chamber with rounded corners is easy to clean and convenient to use;
  • Microcomputer servo controller with digital temperature and humidity display (SD/GSD models have this function);
  • LCD touchscreen controller displays all functions graphically and in text, with curves showing equipment operating status (SDP/GSP models have this function);
  • Continuous operation: two imported hermetic compressor units switch automatically to ensure continuous operation;
  • Continuous operation requires no defrosting, avoiding chamber temperature and humidity fluctuations caused by defrosting during use;
  • An imported humidity sensor avoids frequent wick replacement associated with conventional wet- and dry-bulb sensors;
  • GSD/GSP series provides stepless illuminance adjustment;

Safety Functions

  • Air circulation and heating stop automatically when the door opens to prevent temperature overshoot;
  • Automatic circulating fan speed control prevents sample evaporation caused by excessive airflow;
  • Password-protected screen locking prevents incorrect operation by non-laboratory personnel;
  • An independent temperature-limit alarm system provides audible and visual warnings;
  • Temperature-deviation alarms; compressor overheating, overload and overpressure protection; water-shortage protection;
  • Wired monitoring alarm system (SDP/GSP series has this function)

3Q qualification and calibration services are available.

Tiered Access Management

  • Multiple individual accounts can be assigned according to equipment-management requirements, with three controller access levels: administrator, operator and guest. The touchscreen supports Chinese and English input, allowing users to log in under their actual names. Detailed user-operation logs can also be queried for equipment operation, maintenance and audit trails;

Technical Specifications

Technical Specification / Model Pharmaceutical Stability Test Chamber Combined Pharmaceutical Stability Test Chamber
LHH-500SD
LHH-500SDP
LHH-800SD
LHH-800SDP
LHH-1000SD
LHH-1000SDP
LHH-1500SD
LHH-1500SDP
LHH-500GSD
LHH-500GSP
LHH-800GSD
LHH-800GSP
LHH-1000GSD
LHH-1000GSP
LHH-1500GSP
Temperature Control Range 0~65°C 0~65°C (15~50°C with lighting)
Temperature Fluctuation / Uniformity ±0.5°C/±2°C~±3°C
Humidity Range / Deviation 25~95%RH/±3%RH
Lighting Not provided 0-6000LX adjustable, ≤±500LX
Refrigeration System Two independent original imported hermetic compressors with automatic changeover
Controller Imported programmable touchscreen controller
Sensor Temperature: Pt100 platinum resistance sensor; humidity: capacitive humidity sensor
Ambient Operating Temperature +5~30°C
Power Supply AC220V 50HZ AC380V 50HZ AC220V 50HZ AC220V 50HZ AC380V 50HZ
Rated Power / Operating Power (W) 3750 / approx. 1400 3400 / approx. 2720 3500 / approx. 2800 6000 / approx. 4800 3910 / approx. 1400 3600 / approx. 2880 3700 / 6000 / approx. 2960 / approx. 4800
Capacity 500L 800L 1000L 1500L 500L 800L 1000L/1500L
Chamber Dimensions (mm)
W×D×H
670×725×1020 730×750×1480 730×890×1480 1550×590×1650 670×725×1020 730×750×1480 730×890×1480
1550×590×1650
Overall Dimensions (mm)
W×D×H
850×1100×1930 870×1085×2065 870×1220×2065 2110×890×2000 850×1100×1930 870×1085×2065 870×1220×2065
2110×890×2000
Sample Trays (Standard) 4 shelves
Safety Devices Compressor overheating protection, fan overheating protection, overtemperature protection, compressor overpressure protection, overload protection, water-shortage protection, wired monitoring alarm system
Remarks 1. Built-in printer and RS485 interface standard;
2. SDP and GSP series also include a USB interface and wireless alarm system as standard; optional on SD and GSD series.
3. High-precision digital recorder optional
4. Combined pharmaceutical stability test chambers have an illuminance monitor.
5. GSD/GSP can be fitted with optional two- to three-tier lighting.
6. 800 liters: single door; 1000/1500: double doors.