High-performance portable pH conductivity meter, including 4 measurement functions, the meter is suitable for outdoor measurement of pH, mV, conductivity and total dissolved solids (TDS) of samples.
Features of pH mode:
• High-performance portable pH conductivity meter with backlit LCD display.
• 1 to 5 point calibration, automatic identification of USA/NIST/DIN standard buffers.
• Automatic electrode diagnosis, display pH slope and zero offset.
Conductivity/TDS mode features:
• 1 to 5 point calibration, automatic identification of conductivity standards.
• Optional cell constant (K=0.1/1/10), reference temperature (20/25°C), TDS conversion coefficient, linearity and pure water compensation modes.
Common features:
• Automatic temperature compensation to ensure accurate measurement of the full range.
• Automatic end point lock, keep stable readings for easy browsing and recording.
• Calibration due reminder, reminding the user to calibrate the meter regularly.
• Automatic power off, effectively saving battery power consumption.
• The setting menu can customize the number of calibration points, display resolution, stability conditions, temperature units, date and time, etc.
• Reset function, automatically restore the meter to the factory default settings.
• Expand memory, store or recall up to 500 sets of data.
• USB communication interface, easy to transfer data to a personal computer.
• Multi-mode power supply scheme (battery, power adapter, computer USB port) to ensure the smooth application of the meter.
Technical
| model | Bante901P |
| pH | |
| Measuring range | -2.000~20.000pH |
| display resolution | 0.001, 0.01, 0.1pH, optional |
| measurement accuracy | ± 0.002pH |
| Calibration point | 1 to 5 points |
| pH buffer options | USA, NIST, DIN, 2 custom buffers |
| mV | |
| Measuring range | ± 1999.9mV |
| display resolution | 0.1, 1mV, optional |
| measurement accuracy | ± 0.2mV |
| Conductivity | |
| Measuring range | 0.01~20.00, 200.0, 2000µS/cm, 20.00, 200.0mS/cm |
| display resolution | 0.001, 0.01, 0.1, 1 |
| measurement accuracy | ± 0.5% FS |
| Calibration point | 1 to 5 points |
| Calibration solution | 10µS, 84µS, 1413µS, 12.88mS, 111.8mS |
| Temperature compensation coefficient | Linear (0.0~10.0%/ ° C), pure water |
| Reference temperature | 20/25 ° C |
| Conductivity cell constant | K=0.1, 1, 10, custom |
| TDS | |
| Measuring range | 0.00~10.00, 100.0, 1000ppm, 10.00, 200.0ppt |
| display resolution | 0.01, 0.1, 1 |
| measurement accuracy | ± 1% FS |
| TDS conversion factor | 0.1~1.0 ( default 0.5) |
| temperature | |
| Measuring range | 0~105°C/32~221°F |
| display resolution | 0.1°C/0.1°F |
| measurement accuracy | ±0.5°C/±0.9°F |
| Offset calibration | 1 point, measured value ±10°C |
| General parameters | |
| Temperature compensation range | 0~100°C/32~212°F, manual or automatic |
| Data lock | Manual or automatic end point lock |
| Stability conditions | Standard or high precision |
| Calibration due reminder | 1 to 31 days or closed |
| Automatic shut-down | 10, 20, 30 minutes after no key operation |
| Data storage | 500 groups |
| Communication output | USB |
| Connector | BNC, 6-pin DIN, 3.5mm microphone jack |
| Display screen | Custom LCD display (80×60mm) |
| Power type | 3 AA batteries or DC5V power adapter |
| Dimensions | 170(L)×85(W)×30(H)mm |
| Meter weight | 300g |