Character
- Single-chip microcomputer control, LCD Screen (128*64 Dots)
- Large LCD display can display multiple sets of data
- Huge memory space to store multiple sets of data and curves
- Auto zero,100% adjustment function
- Wavelength automatic adjustment
- Wrap lights, automatic filter change, large sample pool (5mm ~ 100mm)
- Standard samples having a maximum ten-point standard curve measurement functions built
- Direct input by K, B factor for quantitative measurement standard curve
- Can be directly input and the corresponding standard concentration values establish a standard curve for quantitative measurement
- Adjustable electric saving standard curve parameter measuring set, can save a plurality of standard curve
- Can be edited at any time of the standard curve, user-friendly
- General Parallel printer interface with or slightly beat interfaces can print the title bar, the measurement data, curve parameters, the standard curve and the curve sample points
- With USB interface
- Can achieve more accurate and flexible measurement requirements (optional) via PC control
| PRODUCT MODEL: | UV-1200 UV-V |
| OPTICAL SPECIFICATIONS | |
| Wavelength Range | 200 ~ 1000 nm |
| Bandwidth | 2nm |
| Wavelength Accuracy | ±1.0 nm |
| Wavelength Repeatability | 0.2nm |
| Photometric Accuracy | ±0.3% t |
| Photometric Repeatability | 0.15% t |
| Stray Light | ≤0.05% t (220nm Nal, 340nm NaNO2) |
| Stability | 0.001A/h@500nm |
| Photometric Mode | T,A,C,E |
| Wavelength Setting | Automatic |
| Photometric Display Range | -0.3~3A |
| Display Mode | LCD Screen (128*64 Dots) |
| Detector | Import Silicon Photodiode |
| Light Source | Import Deuterium Lamp & Tungsten Halogen Lamp |
| Power Requirement | AC 220V / 50Hz 110V / 60Hz |
| Power | 120W |
| Dimensions (W*D*H) | 460x330x210mm |
| Weight | 11Kg |
To understand the fundamental principles common to all types of spectrophotometers, be sure to read our main article: What Is a Spectrophotometer & How Does It Work? The Ultimate Guide.