BK-1200 Auto Chemistry Analyzer

Features:
①. 150 Sample positions.
②. 160 Reaction cuvettes.
③. 180 Refrigerated reagent positions.
④. Optical filter/fiber bragg grating selectable.
⑤. 800 Tests per hour 1200 Tests with ISE module.

⑥. Built-in sample barcode system optional.

⑦.Probe with anti-collision function, liquid level detection function.

Model BK-1200
Overall
Performance
Throughput 800Tests/hour, 1200Tests/hour with ISE(Optional)
Analysis Method End-point, Fixed-time, Rate(Kinetic), Turbidimetry
Certificates CE, FDA, ISO9001, ISO14001, ISO13485
Sample & Reagent System Sample Positions 150sample positions
Reagent Position 180reagent positions
Sample Volume 2~70ul
Reagent Volume 20~350ul
Probe Teflon coating,withanti-collision function,Liquid level detectionfunction
Probe Washing Automatic washing interior and exterior
Reagent Cooling Refrigerated tray with independent switch
Reaction System Temperature Control Water bath37±0.1℃, real-time monitoring
Cuvettes 160reusable cuvettes, optical length 6mm
Mixer Probe Independent stirring
Washing Automatic cuvettes washing
STAT function YES
Optical System Light Source 12V/20W halogen lamp/fiber bragg grating optional
Spectrophotometry Post-spectral spectrophotometry
Wavelength 12 wavelength from 340~800nm
Absorbance -0.3~3.0Abs
Calibration & QC Calibration Linear:K factor, 1-point ,2-point and multipoint linear

Non-Linear:Spline,Polygon, Index, Ogarithm, Logit-4P, Logit-5P

Quality Control Real-time QC, Westgard multi rule, Cumulative Sum Check,Twin Plot(2D)
Data Management Software Windows 7/8/10 , 32or 64bit
LIS System Bi-direction,support HL7 protocol
Interface LAN port access
  Printer External,multiple reporting mode available
Working Conditions Power Supply AC220V±10% 50/60Hz, 110V±10% 60Hz, 1500W
Temperature 10~30℃
Water Consumption Deionized water:20L/H
Humidity 30-80%
Size & Weight External Size (W*D*H) 1360*791*1157mm
Net Weight 286kg
Package Size (W*D*H) 1718*929*1434mm
Gross Weight 390kg
Categories: , Tag: