LS-50HV/LS-75HV/LS-100HV Vertical Pulse Vacuum Steam Sterilizers
Use
The vertical pressure steam sterilizers are assembled with heating system, micro-computer controlled system, over heat and over pressure protection system, which are reliable for the sterilizing effects, convenient of operation and energy conservation. They are ideal equipments for clinics, scientific research institutes and other organizations to sterilize surgical instruments, fabrics, glasses, culture media etc.
Pulse vacuum automatic
Characteristics
- Microcomputer controlled, digital display of working statues, touch type key.
- It dynamically shows the temperature, pressure time, vacuum rate and the cycle status.
- With fully automatically route of the water inlet, pulse vacuum, heating, sterilizing, draining, exhausting and vacuum drying.
- 3 times pulse vacuum, the vacuum rate would reach -0.081 MPa after that, which ensure the completely extracting of inner air from the dressing and A,B type hollowed appliance.
- With 4 fixed programs for the appliance, medical dressing, rubber and fluid, the user also could set the program as they need randomly.
- Equipped with the PT/TT testing connector, easy for running the testing.
- Multiple pulse vacuum drying after sterilizing, which ensure the good sterilizing result.
- Equipped with 0.22 \mu m bacteria filter, it ensures the air which enters the inner vacuumed chamber is sterile, avoid the second time contamination.
- Hand wheel type for quick-open the door, self-inflating type seal.
- Steam-water inner circulation system: no steam discharge, and the environment for the sterilizing will be clean and dry.
- Automatically shut off with beep reminding after sterilization.
- Fully high quality stainless, equipped with double stainless steel baskets.
- The cycle cannot be started if the door is not sealed. Once the cycle is started and running, the sealed door is locked and cannot be opened to ensure operational safety.
- Build-in printers supports the real-time printing.
- Reservation function: The date, year, month, day, hour and minute can be set.
Technical Data
| Model Technical Data |
LS-50HV | LS-75HV | LS-100HV |
|---|---|---|---|
| Chamber volume | 50L Φ 340 × 550 mm | 75L Φ 400 × 600 mm | 100L Φ 440 × 650 mm |
| Design pressure | 0.3MPa | ||
| Design temperature | 150°C | ||
| Max. working pressure | 0.25MPa | ||
| Max. working temperature | 138°C | ||
| Heat average | ≤ ± 1°C | ||
| Timer range | 0~99min | ||
| Power | (3.2+1.6)KW/AC220V.50Hz | (3.2+1.8)KW/AC220V.50Hz | (3.2+1.9)KW/AC220V.50Hz |
| Overall dimension(mm) | 660 × 460 × 1160 | 700 × 520 × 1250 | 820 × 560 × 1270 |
| Transport dimension(mm) | 740 × 550 × 1270 | 780 × 600 × 1410 | 950 × 680 × 1430 |
| G.W/N.W | 137kg/114kg | 168kg/142kg | 190kg/168kg |
Note: Equipped with seperate steam generator