Introduction
- Single-chip microcomputer control, Touch LCD Screen (7 inches).
- Large LCD display can display multiple sets of
- Huge memory space to store multiple sets of data and curves.
- Auto zero,100% adjustment
- Wavelength automatic
- Wrap lights, automatic filter change , large sample pool (Smm – 100mm).
- Standard samples having a maximum ten-point standard curve measurement functions built.
- Direct input by K, B factor for quantitative measurement standard
- Can be directly input and the corresponding standard concentration values establish a standard curve for quantitative measurement.
- Adjustable electric saving standard curve parameter measuring set, can save a plurality of standard
- Can be edited at any time of the standard curve , user-friendly
- General Parallel printer interface with or slightly beat interfaces can print the title bar, the measurement data , curve parameters , the standard curve and the curve sample points.
- With USB
- PC control can be achieved through more accurate and flexible measurement requirements standard ).
Optical Specifications
PRODUCT MODEL: | UV-1500CPC UV-V |
OPTICAL SPECIFICATIONS | |
Wavelength Range | 190 ~ 1100 nm |
Optical circuit structure | C-T type |
Bandwidth | 2nm |
Wavelength Accuracy | ±0.5 nm |
Wavelength Repeatability | ≤0.2nm |
Photometric Accuracy | ±0.3% t |
Photometric Repeatability | 0.15% t |
Stray Light | ≤0.04% t (220nm Nal, 340nm NaNO2) |
Stability | 0.001A/h@500nm |
Photometric Mode | T,A,C,E |
Wavelength Setting | Automatic |
Photometric Display Range | -0.3~3A |
Display Mode | Touch LCD Screen (7 inches) |
Detector | Import Silicon Photodiode |
Light Source | Import Deuterium Lamp & Tungsten Halogen Lamp |
Power Requirement | AC 220V / 50Hz 110V / 60Hz |
Power | 120W |
Dimensions (W*D*H) | 460x350x220mm |
Weight | 15Kg |