TU1000/TU200 Benchtop Turbidity Meter

TU1000

● 1-7 points calibration using Formazin standards or zero water.
● Autoranging from 0-1000 NTUs.
● Data capacity of up to 2000 sets (GLP-compliant).
● Data stored can be transferred into the computer by USB or RS-232 communication interface.

TU200

● 1-5 points calibration using Formazin standards or zero water.
● Autoranging from 0-200 NTUs.

GENERAL FEATURES

● 880nm LED Light Source.
● Uses Nephelometric principle with 90° detection.
● Meets ISO 7027 and offers consistent light source for higher accuracy.
● Selectable turbidity units, including the NTU, FNU and EBC.
● Settable parameters, including user ID, number of calibration points, data and time, etc.

Model TU1000 TU200
Measuring Unit Range (0 20.00) NTU,
(20.0
200.0) NTU (200~1000) NTU
(0 20.00) NTU,
(20.0
200.0) NTU
Accuracy ≤ ±6% ≤ ±8%
Repeatability ≤0.5% ≤1.0%
Fluctuation ≤ ±0.5% FS/30min ≤ ±1.0% FS/30min
General Power Supply AC Adapter, 200-240 VAC input
Dimension (mm)& Weight (kg) 330×250×130, 2.5