BK-IN 1 Pro

Infant Incubator

The infant incubator provides newborns with a temperature and humidity adjustable environment similar to that of the mother’s womb,which does not apply to the transfer of the baby.

Features:

1.Microprocessor based servo control mode: air temperature, skin temperature, humidity, oxygen concentration.

2.LCD display: set temperature, skin temperature, humidity, oxygen concentration, heat power rate.

3.Double-layer thermostat cover; 4 operating windows and 2 iris ports.

4.Self–testing function;22 kinds of audible and visual alarms.

5.>37℃ preset mode.

6.Independent locking device for front door.

7.The inclination of infant bed is adjustable.

 

Model BK-IN 1 Pro
Specification of Hood
Hood Size (W*D*H)mm 865*455*490
 Mattress Size(W*D)mm 770*380
Thermostat Cover Double-layer
Access Port 6(2 Iris Ports)
Operating Window 4
Infusion Seal 8
Technical Parameters
Power Supply AC110/220V±10%, 50/60Hz
Consumption 1320W
Heater Consumption Heating chamber: 290W; Water box:170W
Display LCD
Control Model Air, Skin, Humidity, Oxygen concentration
Heating Indicator 0~100%, 10% increments
Air Temp. Control Range(℃) 25.0~37.0; 37.0~38.0
Skin Temp. Control Range(℃) 34.0~37.0; 37.0~38.0
Skin Sensor Temp. Display Range(℃) 5.0~65.0
Skin Sensor Temp. Accuracy(℃) ±0.3
Temp. Rise Time(Ambient Temp. 22℃±2℃) ≤50 min
Temp. Fluctuation(℃) ≤0.5
Mattress(Horizontal)Temp. Uniformity(℃) ≤0.8
Mattress(Tilted)Temp. Uniformity(℃) ≤1.0
The Difference between the Average Air Temp. and the Set Temp.(℃) ≤1.5
Water Box Capacity(ml) 1200
Relative Humidity Display Range 0%~99% RH
Relative Humidity Control Range CAH%RH~90%RH
Humidity Display Accuracy ±10%
Oxygen Concentration Display Range 0%~99%
Oxygen Concentration Display Accuracy ±5%
Oxygen Sensor Lifetime 10000 hours under 100% oxygen concentration
Noise Level of Inside Incubator ≤55dB[A]
Infant Bed Tilt Angle ±10%
Operating Condition
Working Temp. Range(℃) 20~30
Transport and Storage Temp. Range(℃) -20~55
Working Relative Humidity 30%~75% RH
Transport and Storage Relative Humidity ≤93%RH
Ambient Air Flow Rate <0.3m/s
Working Atmospheric Pressure 800hPa~1060 hPa
Transport and Storage Atmospheric Pressure 500 hPa~1060 hPa
Application Environment Elevation ≤2000m
Overvoltage Category
Pollution Grade 2
General Alarm Power Failure, Fan Failure, Temp.Deviation, Sensor failure,Over Temperature,Water Shortage, etc
System Alarm Communication Deviation, Temp. and Humidity Deviation
Standard Accessory RS232 interface connector, X ray cassette, IV pole, tray, drawer*3.
External Size(W*D*H)mm 1050*675*1400~1600
Net Weight 117kg
PackageSize(W*D*H)mm

Control Unit and Hood

Base

 

1190*780*890

1100*760*830

Gross Weight 127kg

 

Category: Tag: