F96S Fluorescence Spectrophotometer
The F96S Fluorescence Spectrophotometer is ideal for quantitative fluorescence measurements in the UV and visible light regions. By selecting a special photomultiplier tube (PMT) detector, its detection range can be extended into the red-sensitive spectrum. It supports two operational modes: standalone and PC-controlled. In standalone mode, the built-in microcomputer system provides functions such as fluorescence intensity measurement, standard curve creation and storage, concentration measurement, data printing, auto-zero, and automatic background subtraction.
The F96S Fluorescence Spectrophotometer is a high-precision, multi-functional instrument primarily used for measuring and analyzing the fluorescence signals of samples.
This instrument incorporates advanced technology and design principles to deliver accurate and reliable fluorescence measurement results. It is equipped with a high-sensitivity detector and a highly stable light source, enabling measurements across a broad wavelength range and supporting a variety of fluorescence analysis applications.
Key Features of the F96S Fluorescence Spectrophotometer:
- Wide Wavelength Range: The instrument can perform wavelength scans within the 200-900nm range, providing high-quality fluorescence emission spectra.
- High Sensitivity: Through optimized design and advanced detector technology, the instrument can accurately detect low-concentration samples and weak fluorescence signals.
- Multi-Functional Operation: Users can select different operational modes according to their experimental needs, such as single-point measurement and kinetic scans. It also supports custom programming to meet special requirements.
- Data Processing and Export: The F96S is equipped with powerful and easy-to-use data processing software that allows for data analysis, peak identification, and data export.
- User-Friendly Design: The instrument features an intuitive interface and an easy-to-use control panel, allowing users to quickly set up experimental parameters and obtain accurate results.
| technical parameter | technical indicators |
| instrument | F96S |
| detection sensitivity | Water Raman peak signal-to-noise ratio:S/N ≥90(P-P) |
| Excitation wavelength bandwidth | 10/20nm |
| Emission wavelength bandwidth | 10nm |
| Emission wavelength accuracy | ±1.0nm |
| Emission wavelength repeatability | ≤0.5nm |
| light source | High-intensity LED cold light source |
| Optical system | C-T Monochromator Grating Automatic Wavelength Scanning |
| Fluorescence receiver | Binsong original photomultiplier tube |
| Excitation wavelength selectable range (LED light source) | 250~600nm |
| Emission wavelength range | (200-650)nm, expandable to 200-900nm |
| Wavelength scanning speed | Fastest 30000nm/min, fine scan 15nm/min |
| measure linearity | Correlation coefficient≥0.995 |
| Gain adjustment range | 1~17 gears optional |
| Integral response time | 6 gears optional:0.1s~4s |
| time scan settings | Can be set arbitrarily, up to 640px00 seconds |
| Data output method | Microsoft (R) Excel format, bmp picture format |