Mold Incubator
In line with worldwide environmental trends, well-known-brand compressors and circulating fans provide high efficiency, low energy consumption, energy savings, long service life and reduced noise. Cooling time is reduced by more than 40% compared with conventional low-temperature equipment.
Application Overview
Suitable for research, educational and production departments in environmental protection, health and epidemic prevention, pharmaceutical testing, agriculture, animal husbandry and aquaculture. This dedicated constant-temperature equipment is used for water analysis and BOD determination, culture and storage of bacteria, mold and microorganisms, plant cultivation and breeding tests.
Product Features
- A mirror-finished stainless-steel chamber has rounded corners for easy cleaning and adjustable shelf spacing;
- A well-known-brand compressor with environmentally friendly refrigerant provides high efficiency and low energy consumption to promote energy savings;
- Integral foam insulation provides effective thermal insulation, low energy loss and more precise temperature control, improving insulation efficiency by more than 25% (70/150/250 series);
- A microcomputer PID temperature controller provides precise, reliable control with low fluctuation and a timer;
- Three circulating fan speed settings prevent sample evaporation caused by excessive airflow during experiments;
- An independent temperature-limit alarm system automatically interrupts operation when the limit is exceeded, ensuring safe experiments without accidents (optional);
- An optional printer or RS485 interface connects to printers and computers to record changes in temperature parameters;
- LCD display with menu-driven operation (available on models with “F” and Type II);
Technical Specifications
| Technical Specification / Model | Biochemical Incubator | Mold Incubator | Mold Incubator (Humidifying Type) |
|||||
|---|---|---|---|---|---|---|---|---|
| LRH-70 LRH-70F |
LRH-150 LRH-150F |
LRH-250 LRH-250F |
LRH-500F | MJ-70-I MJ-70F-I |
MJ-150-I MJ-150F-I |
MJ-250-I MJ-250F-I MJ-500F-I |
MJ-150-II MJ-250-II MJ-500-II |
|
| Temperature Control Range | 0~60°C | 0~60°C | 0~60°C | With humidification: 15~45°C Without humidification: 10~45°C |
||||
| Temperature Resolution | 0.1°C | |||||||
| Temperature Fluctuation | High temperature ±0.5°C; low temperature ±1.0°C | |||||||
| Temperature Uniformity | ±1.5°C (test point: 25°C) | ±2.5°C (test point: 25°C) |
±1.5°C~±2.5°C (test point: 25°C) | |||||
| Humidity Control Range | —— | 60~85%RH | ||||||
| Humidity Deviation | —— | ±5~8%RH | ||||||
| Chamber Insulation | Integral foam insulation | —— | Integral foam insulation | Integral foam insulation / —— | Integral foam insulation / —— | |||
| Supply Voltage | AC220V 50HZ | |||||||
| Ambient Operating Temperature | +5~30°C | |||||||
| Input Power | 450W | 500W | 600W | 2100W | 450W | 500W | 600W 2250W |
680W 850W 2250W |
| Chamber Dimensions (mm) W×D×H | 400×350×500 | 503×370×808 | 540×460×1000 | 670×720×1020 | 400×350×500 | 503×370×808 | 540×460×1000 670×720×1020 |
550×405×670 600×500×830 670×720×1020 |
| Overall Dimensions (mm) W×D×H | 530×560×1080 | 600×630×1360 | 637×662×1590 | 850×1100×1930 | 530×560×1080 | 600×630×1360 | 637×662×1590 850×1100×1930 |
690×800×1430 750×900×1580 850×1100×1930 |
| Shelves (Standard) | 2 shelves | 3 shelves | 2 shelves | 3 shelves | ||||
| Timer Range | 0~5999min | |||||||
| Remarks | In addition to the features of Type I, Type II mold incubators provide microcomputer humidity control. ※ Performance parameters are tested without a load at an ambient temperature of 20°C and ambient humidity of 50%RH. |
|||||||
Notes: 1. Models with “F” and Type II use an LCD controller; models without “F” use a digital-display controller.
2. LRH-500F delivery time is 15 days. 3. Type II mold incubators include a Φ25mm test port on the left side as standard.
Optional Accessories:
1. Multi-segment programmable temperature/humidity controller (Type II mold incubators) or multi-segment programmable temperature controller (biochemical and Type I mold incubators)
2. Built-in printer
3. Paperless recorder
4. USB flash drive data storage
5. Independent temperature-limit controller
6. Left-side 25/50mm-diameter test port
7. Left-side 100mm-diameter test port 8. BOD socket
9. RS485/232 interface and communication software
Note: choose one of USB flash drive, RS485/232 or printer.