BK6400 Auto 5 parts Hematology analyzers are used to run tests on blood samples. They are used in the medical field to do white blood cell counts, complete blood counts, reticulocyte analysis.
Features:
①.60 Tests/hour .
②.Laser Scatter Technology.
③.Constant linear sheath flow.
④.2 test modes & 2 sample modes.
⑤.3-D topographic maps for WBC.
⑥.Integral titanium incubation system.
⑦.28 parameters, 2 histograms, 2 scattergrams.
Model | BK-6400 | |||
Throughput | 60Tests/hour | |||
Assay Items | 5 parts, 28 parameters, 2 histograms, 2 scattergrams | |||
Test mode | CBC+5DIFF, CBC | |||
Measuring Principle | Laser Scatter Technology | |||
Testing Items | Regular Parameters: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, PDW, MPV, PCT, P-LCR, NEU#, NEU%, LYM#, LYM%, MON#, MON%, EOS#, EOS%, BAS#, BAS%
Research Parameters: ALY#, ALY%, IG#, IG% |
|||
Performance | Item | Linear range | Carry over | CV |
WBC | 1.0-99.9×10^9/L | ≤0.5% | ≤2.0% | |
RBC | 0.3-7.0×10^12/L | ≤0.5% | ≤1.5% | |
HGB | 20-240g/L | ≤0.5% | ≤1.5% | |
PLT | 20-999×10^9/L | ≤1.0% | ≤4.0% | |
Storage | 100,000 results storage including histogram and scattergram | |||
LIS Interface | Support HL7 protocol | |||
QC and Calibration | Multiple QC rules, including L-J, X-B etc
Auto and manual calibration function for both whole blood mode and pre-diluted blood mode |
|||
Incubation | Integral titanium incubation system | |||
Operation Environment | Temperature: 15~30℃ | |||
Humidity: 30%~85% | ||||
Power supply | AC220V±10% 50/60Hz; 110V±10% 60Hz; 600VA | |||
Package Size | 730*610*830mm(W*D*H) | |||
Gross Weight | 80kg | |||
ReagentPackage Size | 320*320*300mm (W*D*H);340*290*280mm (W*D*H) | |||
ReagentGross Weight | 24kg |