Description

3 part diff of WBC. ②. 20 parameters+3 histograms. ③. 60 tests per hour.④.touch screen

Features:

①.Throughput: 60T/H

②.8 inch touch screen

③.20 parameters + 3 histograms

④.3 counting modes

⑤.100,000 sample results

⑥.Support LIS and external printer

ModelBK-6190
Throughput60Tests/hour
Measuring PrincipleImpedance for cell counting Cyanide-free method for HGB
Aperture DiameterWBC chamber:100μm,RBC/PLT chamber:70μm
ParametersWBC, Neu#, Lym#, Mid#, Neu%,Lym%, Mid%, RBC, HGB, HCT, MCV, MCH, MCHC,RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, P-LCR
Histograms3 histograms(includingWBC, RBC, PLT histogram)
Sample VolumeNormal:10μl; Predileted:20μl
CalibrationManual & auto calibration
Counting ModesVenous, capilary, prediluted
PrinterBuilt-in thermal printer,USB to external printer(optional)
Data Input8 inch touch screen, keyboard, mouse
OutputRS232 port, USB port, VGA port. LIS system avaliable
Storage100,000 results with histogram
Carry-over rateParameterCV
WBC≤ 2%
RBC≤ 1%
HGB≤ 2%
PLT≤ 2%
BackgroundParameterBackground
WBC≤ 0.2×10^9/L
RBC≤ 0.02×10^12/L
HGB≤ 1 g/L
HCT≤ 0.5%
PLT≤ 10×10^9/L
 

Precision

 

ParameterCVMeasurement range
WBC≤ 2.0%(4.0-15.0×10^9)/L
RBC≤ 1.9%(3.5-6.0×10^12)/L
HGB≤ 1.9%(110.0-180.0g)/L
MCV≤ 0.4%(80.0-110.0)fL
PLT≤ 4.0%(100.0-500.0×10^9)/L
LinearityParameterMeasurement rangeCV
WBC(0-5.9×10^9)/L

(6.0-99.9×10^9)/L

(±0.3×10^9)/L

±5%

RBC(0-0.99×10^12)/L

(1.0-9.99×10^12)/L

(±0.05×10^12)/L

±5%

HGB(0-99.0g)/L

(100.0-300.0g)/L

(±2.0g)/L

±2%

PLT(0-99×10^9)/L

100-999×10^9)/L

(±8.0×10^9)/L

±10%

Power supplyAC 220±10%,60/50Hz, 110±10%,60Hz
Package Size580*470*590mm (W*D*H)ReagentPackage Size450*410*310mm (W*D*H)