Comparative Evaluation of China and Imported 5mL Pipettes

  1. Introduction

Pipettes are essential tools in modern laboratories. Beyond chemical analysis, they are extensively used in biology, medicine, environmental science, and food safety labs, with a massive volume of consumption. As an instrument exporter, HINOTEK offers various types of pipettes.

Currently, imported pipettes are expensive; brands like Gilson, Eppendorf, and Rainin generally cost between 1,000 to 2,000 RMB per unit. The most prominent domestic brand is DLAB, priced at approximately 1/10th of imported brands. Additionally, there are numerous second-tier domestic brands priced under 100 RMB, which are exceptionally affordable. While the mechanical principle of a pipette is simple and has low barriers to entry regarding materials and manufacturing, laboratory supplies demand stability and reliability. This requires precision craftsmanship and data accuracy rather than a sole focus on low cost.

This study presents a comparative evaluation of 5mL pipettes from Gilson, DLAB, and six other domestic brands. The primary objectives are:

  1. To provide an objective understanding of the quality of domestic brands, moving beyond the “cheap goods” stereotype.
  2. To clarify evaluation methodologies so users can “purchase with insight and use with clarity.”
  1. Test Subjects

All pipettes selected for this evaluation have a 5mL volume range.

  • Imported Representative: Gilson (France).
  • Domestic Tier 1 Representative: DLAB (Beijing). (These were selected simply because they were available in our laboratory.)
  • Other Domestic Brands: Six brands were randomly selected based on ease of purchase via JD.com (all priced under 100 RMB).

Note: The selection was entirely random and objective, intended to represent the overall level of domestic pipettes rather than targeting specific brands. The author has no conflict of interest with any of these brands. Unauthorized use of this content for commercial promotion is strictly prohibited.

  1. Sensory Evaluation

(1) Appearance

The seven domestic pipettes are shown in Figure 1. Their appearance and dimensions are nearly identical, with differences limited to the printed logos, color schemes, and minor details. It is suspected they may originate from the same OEM with different branding. Craftsmanship is generally consistent, with only slight differences compared to the imported brand. Only the “Fenyi Sihai” brand performed slightly worse, specifically regarding the tip cone (where the tip is attached), which was not smooth and exhibited surface roughness and burrs (Figure 2). This results in less smooth tip mounting and poses a risk of poor sealing. Other brands featured very smooth, mirror-like tip cones, allowing for smooth attachment and ejection.

(2) Tactile Feel (Hand-feel)

Evaluation was based on grip, pressing force, and rebound.

  • Grip: Due to the nearly identical shapes, the grip feel is similar across the domestic brands. The only difference lies in the finger hook (where the index finger rests), as shown in Figure 3. The Gilson design is highly ergonomic; the surface is perfectly smooth and contours to the index finger joint, preventing pressure points during use. Domestic models lag significantly here; although they feature rounded edges to varying degrees, they do not fit the finger shape as well, leading to a noticeable “digging” sensation that can cause pain during prolonged use.
  • Pressing Force: Ideally, the force should be moderate, with a distinct difference between the first and second stops (blow-out). The Gilson excels here, requiring minimal effort and reducing hand fatigue. Among domestic brands, only BKMAM performed well, approaching the feel of the imported model. Other brands had suboptimal spring tension—the first stop was too stiff (requiring excessive force), while the second stop was too weak, making the intermediate stop point indistinct.
  • Rebound: The piston movement should be smooth with moderate return force and speed. BKMAM provided the best experience, feeling even smoother than the imported model. The other five domestic brands suffered from pistons that were too tight or had insufficient return force. Insufficient rebound force not only affects tactile feel but also impacts accuracy and durability (discussed later). Conversely, the Fenyi Sihai model had a spring that was too strong combined with a piston that was too loose, causing a violent rebound that could lead to liquid surging into the internal shaft of the pipette.

Table 1: Summary of Sensory Evaluation

Brand Craftsmanship Grip Pressing Rebound
Gilson (France) ★★★★★ ★★★★★ ★★★★★ ★★★★
DLAB (Beijing) ★★★★ ★★★ ★★★ ★★★
BKMAM ★★★★ ★★★ ★★★★ ★★★★★
JOANLAB ★★★★ ★★ ★★★ ★★★
LICHEN ★★★★ ★★ ★★★ ★★★
LABSHARK ★★★★ ★★ ★★★ ★★★
ONiLAB ★★★★ ★★ ★★★ ★★★
Fenyi Sihai ★★★ ★★ ★★
  1. Accuracy Evaluation

4.1 Methodology

Volume accuracy was determined via the Gravimetric Method. A 3-decimal place balance 0.001 precision) was used with deionized water as the working medium. Weighing was conducted at room temperature, and the actual volume was calculated based on the density of pure water 0.9972 at room temperature).

Prior to testing, the full-scale volume was calibrated to within 5.00±0.02mL. The accuracy tests included:

  • a. Linearity: Measured at three points: Full scale (5ml), half scale(2.5ml), and 10% scale (0.55ml).
  • b. Repeatability: 5 replicates per volume setting to calculate the Standard Deviation (SD).
  • c. Adjustment Reproducibility: After completing a test set, the dial was moved to other volumes and then reset to the target volume after an interval. Three sets were measured to calculate the SD of the mean values between sets. (This is a non-standard metric we call “Adjustment Reproducibility”).

Technical Notes:

  1. The scale interval for domestic 5mL pipettes is 0.05ml The balance precision  0.01g verification, 0.002g repeatability error is significantly higher than the pipette’s precision, meeting the requirements.
  2. The apparent density of water at  20C is 0.9972/mL (buoyancy corrected). Within the standard working temperature range 10C to 30C,  the density varies by only +0.1%~-0.2%,  which is negligible compared to the pipette’s accuracy class.
  3. The plastic material has a coefficient of thermal expansion of approx. 50~100ppm/℃. Even at 100ppm/℃,  the volume expansion is roughly 0.03%/℃, which is negligible.

4.2 Evaluation Results

Table 2 Gilson result:

Group 1 Group 2 Group 3
5mL 5-1 5.001 5-1 4.994 5-1 4.995 Systematic error:-0.002mL

Pipetting repeatability:±0.004mL

Adjustment reproducibility:±0.003mL

5-2 4.991 5-2 4.995 5-2 4.997
5-3 4.994 5-3 4.999 5-3 5.002
5-4 4.995 5-4 5.006 5-4 5.001
5-5 4.993 5-5 5.003 5-5 4.999
Average 4.9948 Average 4.9994 Average 4.9988
Standard deviation 0.0038 Standard deviation 0.0051 Standard deviation 0.0029
2.5mL 2.5-1 2.488 2.5-1 2.492 2.5-1 2.498 Systematic error:-0.003mL

Pipetting repeatability:±0.004mL

Adjustment reproducibility:±0.003mL

2.5-2 2.493 2.5-2 2.502 2.5-2 2.497
2.5-3 2.496 2.5-3 2.500 2.5-3 2.493
2.5-4 2.497 2.5-4 2.499 2.5-4 2.501
2.5-5 2.493 2.5-5 2.499 2.5-5 2.500
Average 2.4934 Average 2.4984 Average 2.4978
Standard deviation 0.0035 Standard deviation 0.0038 Standard deviation 0.0031
0.5mL

 

0.5-1 0.500 0.5-1 0.509 0.5-1 0.511 Systematic error:+0.010mL

Pipetting repeatability:±0.002mL

Adjustment reproducibility:±0.003mL

 

0.5-2 0.507 0.5-2 0.509 0.5-2 0.511
0.5-3 0.505 0.5-3 0.512 0.5-3 0.512
0.5-4 0.508 0.5-4 0.513 0.5-4 0.509
0.5-5 0.508 0.5-5 0.512 0.5-5 0.513
Average 0.5060 Average 0.5114 Average 0.5116
Standard deviation 0.0034 Standard deviation 0.0019 Standard deviation 0.0015

Table 3 DLAB result:

Group 1 Group 2 Group 3
5mL 5-1 5.000 5-1 5.009 5-1 5.010 Systematic error:+0.004mL

Pipetting repeatability:±0.004mL

Adjustment reproducibility:±0.009mL

5-2 4.990 5-2 5.009 5-2 5.016
5-3 4.990 5-3 5.000 5-3 5.010
5-4 4.989 5-4 5.007 5-4 5.014
5-5 5.001 5-5 5.003 5-5 5.012
Average 4.9940 Average 5.0056 Average 5.0124
Standard deviation 0.0060 Standard deviation 0.0040 Standard deviation 0.0026
2.5mL 2.5-1 2.478 2.5-1 2.490 2.5-1 2.500 Systematic error:-0.008mL

Pipetting repeatability:±0.004mL

Adjustment reproducibility:±0.011mL

2.5-2 2.477 2.5-2 2.503 2.5-2 2.499
2.5-3 2.479 2.5-3 2.494 2.5-3 2.504
2.5-4 2.485 2.5-4 2.492 2.5-4 2.499
2.5-5 2.481 2.5-5 2.491 2.5-5 2.504
Average 2.4799 Average 2.4940 Average 2.5012
Standard deviation 0.0032 Standard deviation 0.0053 Standard deviation 0.0026
0.5mL 0.5-1 0.503 0.5-1 0.504 0.5-1 0.502 Systematic error:+0.004mL

Pipetting repeatability:±0.004mL

Adjustment reproducibility:±0.002mL

0.5-2 0.499 0.5-2 0.499 0.5-2 0.503
0.5-3 0.513 0.5-3 0.506 0.5-3 0.504
0.5-4 0.498 0.5-4 0.505 0.5-4 0.500
0.5-5 0.512 0.5-5 0.503 0.5-5 0.501
Average 0.5054 Average 0.5038 Average 0.5024
Standard deviation 0.0071 Standard deviation 0.0027 Standard deviation 0.0016

Table 4 BKMAM Result:

Group 1 Group 2 Group 3
5mL 5-1 5.007 5-1 5.025 5-1 5.013 Systematic error:+0.013mL

Pipetting repeatability:±0.005mL

Adjustment reproducibility:±0.012mL

5-2 5.000 5-2 5.027 5-2 5.009
5-3 4.994 5-3 5.028 5-3 5.018
5-4 5.007 5-4 5.018 5-4 5.005
5-5 4.997 5-5 5.024 5-5 5.015
Average 5.0010 Average 5.0245 Average 5.0120
Standard deviation 0.0059 Standard deviation 0.0039 Standard deviation 0.0051
2.5mL 2.5-1 2.471 2.5-1 2.476 2.5-1 2.476 Systematic error:-0.029mL

Pipetting repeatability:±0.003mL

Adjustment reproducibility:±0.002mL

2.5-2 2.466 2.5-2 2.474 2.5-2 2.470
2.5-3 2.470 2.5-3 2.465 2.5-3 2.470
2.5-4 2.465 2.5-4 2.475 2.5-4 2.470
2.5-5 2.470 2.5-5 2.473 2.5-5 2.473
Average 2.4683 Average 2.4725 Average 2.4717
Standard deviation 0.0027 Standard deviation 0.0044 Standard deviation 0.0027
0.5mL 0.5-1 0.448 0.5-1 0.452 0.5-1 0.450 Systematic error:-0.049mL

Pipetting repeatability:±0.002mL

Adjustment reproducibility:±0.000mL

0.5-2 0.449 0.5-2 0.450 0.5-2 0.451
0.5-3 0.454 0.5-3 0.451 0.5-3 0.453
0.5-4 0.453 0.5-4 0.450 0.5-4 0.452
0.5-5 0.453 0.5-5 0.452 0.5-5 0.450
Average 0.4517 Average 0.4513 Average 0.4515
Standard deviation 0.0027 Standard deviation 0.0010 Standard deviation 0.0013

Table 5 JOANLAB Result:

Group 1 Group 2 Group 3
5mL 5-1 5.005 5-1 5.019 5-1 5.007 Systematic error:+0.008mL

Pipetting repeatability:±0.004mL

Adjustment reproducibility:±0.010mL

5-2 5.007 5-2 5.021 5-2 5.000
5-3 5.011 5-3 5.023 5-3 4.994
5-4 5.002 5-4 5.020 5-4 5.000
5-5 5.006 5-5 5.015 5-5 4.995
Average 5.0062 Average 5.0197 Average 4.9992
Standard deviation 0.0033 Standard deviation 0.0030 Standard deviation 0.0052
2.5mL 2.5-1 2.499 2.5-1 2.511 2.5-1 2.495 Systematic error: 0.000mL

Pipetting repeatability:±0.002mL

Adjustment reproducibility:±0.007mL

2.5-2 2.497 2.5-2 2.507 2.5-2 2.491
2.5-3 2.500 2.5-3 2.506 2.5-3 2.494
2.5-4 2.495 2.5-4 2.508 2.5-4 2.494
2.5-5 2.499 2.5-5 2.506 2.5-5 2.494
Average 2.4980 Average 2.5076 Average 2.4936
Standard deviation 0.0020 Standard deviation 0.0021 Standard deviation 0.0015
0.5mL 0.5-1 0.502 0.5-1 0.504 0.5-1 0.500 Systematic error:+0.002mL

Pipetting repeatability:±0.001mL

Adjustment reproducibility:±0.002mL

0.5-2 0.501 0.5-2 0.502 0.5-2 0.500
0.5-3 0.502 0.5-3 0.504 0.5-3 0.499
0.5-4 0.502 0.5-4 0.502 0.5-4 0.501
0.5-5 0.501 0.5-5 0.501 0.5-5 0.497
Average 0.5020 Average 0.5030 Average 0.4998
Standard deviation 0.0034 Standard deviation 0.0019 Standard deviation 0.0015

 

 

Table 6 LICHEN Result:

Group 1 Group 2 Group 3
5mL 5-1 4.998 5-1 5.016 5-1 5.000 Systematic error:-0.002mL

Pipetting repeatability:±0.006mL

Adjustment reproducibility:±0.012mL

5-2 4.989 5-2 5.013 5-2 4.983
5-3 4.993 5-3 5.011 5-3 4.978
5-4 4.990 5-4 5.004 5-4 4.989
5-5 4.998 5-5 5.016 5-5 4.996
Average 4.9936 Average 5.0120 Average 4.9892
Standard deviation 0.0043 Standard deviation 0.0050 Standard deviation 0.0091
2.5mL 2.5-1 2.491 2.5-1 2.504 2.5-1 2.479 Systematic error:-0.007mL

Pipetting repeatability:±0.004mL

Adjustment reproducibility:±0.013mL

2.5-2 2.491 2.5-2 2.499 2.5-2 2.477
2.5-3 2.491 2.5-3 2.512 2.5-3 2.480
2.5-4 2.496 2.5-4 2.507 2.5-4 2.480
2.5-5 2.502 2.5-5 2.508 2.5-5 2.481
Average 2.4942 Average 2.5060 Average 2.4793
Standard deviation 0.0049 Standard deviation 0.0049 Standard deviation 0.0015
0.5mL 0.5-1 0.504 0.5-1 0.513 0.5-1 0.494 Systematic error:+0.008mL

Pipetting repeatability:±0.004mL

Adjustment reproducibility:±0.006mL

0.5-2 0.510 0.5-2 0.516 0.5-2 0.502
0.5-3 0.507 0.5-3 0.509 0.5-3 0.505
0.5-4 0.516 0.5-4 0.513 0.5-4 0.500
0.5-5 0.503 0.5-5 0.511 0.5-5 0.506
Average 0.5084 Average 0.5128 Average 0.5018
Standard deviation 0.0053 Standard deviation 0.0026 Standard deviation 0.0048

Table 7  LABSHARK Result:

Group 1 Group 2 Group 3
5mL 5-1 4.993 5-1 5.010 5-1 4.999 Systematic error: 0.000mL

Pipetting repeatability:±0.006mL

Adjustment reproducibility:±0.011mL

5-2 4.987 5-2 4.996 5-2 5.013
5-3 4.986 5-3 5.009 5-3 5.015
5-4 4.994 5-4 5.001 5-4 5.009
5-5 4.983 5-5 4.996 5-5 5.010
Average 4.9886 Average 5.002 Average 5.0092
Standard deviation 0.0047 Standard deviation 0.0068 Standard deviation 0.0062
2.5mL 2.5-1 2.504 2.5-1 2.505 2.5-1 2.477 Systematic error:-0.007mL

Pipetting repeatability:±0.004mL

Adjustment reproducibility:±0.016mL

2.5-2 2.500 2.5-2 2.496 2.5-2 2.473
2.5-3 2.500 2.5-3 2.501 2.5-3 2.473
2.5-4 2.510 2.5-4 2.503 2.5-4 2.473
2.5-5 2.496 2.5-5 2.503 2.5-5 2.475
Average 2.5020 Average 2.5016 Average 2.4741
Standard deviation 0.0053 Standard deviation 0.0034 Standard deviation 0.0018
0.5mL 0.5-1 0.508 0.5-1 0.509 0.5-1 0.494 Systematic error:+0.006mL

Pipetting repeatability:±0.002mL

Adjustment reproducibility:±0.006mL

0.5-2 0.509 0.5-2 0.508 0.5-2 0.502
0.5-3 0.507 0.5-3 0.512 0.5-3 0.498
0.5-4 0.508 0.5-4 0.508 0.5-4 0.500
0.5-5 0.509 0.5-5 0.507 0.5-5 0.499
Average 0.5086 Average 0.5092 Average 0.4990
Standard deviation 0.0008 Standard deviation 0.0019 Standard deviation 0.0030

Table 8  ONiLAB Result:

Group 1 Group 2 Group 3
5mL 5-1 5.006 5-1 5.027 5-1 5.005 Systematic error:+0.012mL

Pipetting repeatability:±0.004mL

Adjustment reproducibility:±0.013mL

5-2 5.008 5-2 5.023 5-2 4.995
5-3 5.011 5-3 5.030 5-3 5.00
5-4 5.009 5-4 5.025 5-4 4.994
5-5 5.009 5-5 5.025 5-5 4.99
Average 5.0086 Average 5.0261 Average 5.0002
Standard deviation 0.0018 Standard deviation 0.0027 Standard deviation 0.0062
2.5mL 2.5-1 2.515 2.5-1 2.524 2.5-1 2.479 Systematic error:+0.004mL

Pipetting repeatability:±0.002mL

Adjustment reproducibility:±0.023mL

2.5-2 2.516 2.5-2 2.518 2.5-2 2.475
2.5-3 2.512 2.5-3 2.518 2.5-3 2.480
2.5-4 2.511 2.5-4 2.522 2.5-4 2.480
2.5-5 2.513 2.5-5 2.524 2.5-5 2.480
Average 2.5134 Average 2.5213 Average 2.4787
Standard deviation 0.002 Standard deviation 0.0030 Standard deviation 0.0022
0.5mL 0.5-1 0.528 0.5-1 0.524 0.5-1 0.514 Systematic error:+0.022mL

Pipetting repeatability:±0.002mL

Adjustment reproducibility:±0.008mL

0.5-2 0.527 0.5-2 0.525 0.5-2 0.514
0.5-3 0.526 0.5-3 0.525 0.5-3 0.512
0.5-4 0.531 0.5-4 0.526 0.5-4 0.515
0.5-5 0.524 0.5-5 0.525 0.5-5 0.51
Average 0.5277 Average 0.5255 Average 0.5134
Standard deviation 0.0026 Standard deviation 0.000 Standard deviation 0.0020

Table 9  Fenyi Sihai Result:

Group 1 Group 2 Group 3
5mL 5-1 4.994 5-1 4.982 5-1 4.992 Systematic error:-0.008mL

Pipetting repeatability:±0.004mL

Adjustment reproducibility:±0.001mL

5-2 4.994 5-2 4.996 5-2 4.996
5-3 4.990 5-3 4.989 5-3 4.989
5-4 4.995 5-4 4.994 5-4 4.991
5-5 4.990 5-5 4.996 5-5 4.988
Average 4.9926 Average 4.9914 Average 4.9912
Standard deviation 0.0024 Standard deviation 0.0060 Standard deviation 0.0031
2.5mL 2.5-1 2.503 2.5-1 2.499 2.5-1 2.496 Systematic error:-0.001mL

Pipetting repeatability:±0.002mL

Adjustment reproducibility:±0.003mL

2.5-2 2.504 2.5-2 2.503 2.5-2 2.495
2.5-3 2.498 2.5-3 2.502 2.5-3 2.495
2.5-4 2.502 2.5-4 2.499 2.5-4 2.496
2.5-5 2.499 2.5-5 2.500 2.5-5 2.497
Average 2.5012 Average 2.5006 Average 2.4958
Standard deviation 0.0026 Standard deviation 0.0018 Standard deviation 0.0008
0.5mL 0.5-1 0.506 0.5-1 0.505 0.5-1 0.503 Systematic error:+0.003mL

Pipetting repeatability:±0.002mL

Adjustment reproducibility:±0.002mL

0.5-2 0.502 0.5-2 0.502 0.5-2 0.500
0.5-3 0.502 0.5-3 0.508 0.5-3 0.500
0.5-4 0.500 0.5-4 0.505 0.5-4 0.502
0.5-5 0.507 0.5-5 0.502 0.5-5 0.500
Average 0.5038 Average 0.5048 Average 0.5014
Standard deviation 0.0030 Standard deviation 0.0025 Standard deviation 0.0014

Table 10 Summary of Accuracy Evaluation Results

Gilson Dlab BKMAM JOANLAB LICHEN LABSHARK ONiLAB Fenyi Sihai
Full Scale(5mL) Systematic error -0.002mL +0.004mL +0.013mL +0.008mL -0.002mL 0.000mL +0.012mL -0.008mL
Pipetting repeatability ±0.004mL ±0.004mL ±0.005mL ±0.004mL ±0.006mL ±0.006mL 0.004mL ±0.004mL
Adjustment reproducibility ±0.003mL ±0.009mL ±0.012mL ±0.010mL ±0.012mL ±0.011mL ±0.013mL ±0.001mL
50% Scale

(2.5mL)

Systematic error -0.003mL -0.008mL -0.029mL 0.000mL -0.007mL -0.007mL +0.004mL -0.001mL
Pipetting repeatability ±0.004mL ±0.004mL ±0.003mL ±0.002mL ±0.004mL ±0.004mL ±0.002mL ±0.002mL
Adjustment reproducibility ±0.003mL ±0.011mL ±0.002mL ±0.007mL ±0.013mL ±0.016mL ±0.023mL ±0.003mL
10% Scale

(0.5mL)

Systematic error +0.010mL +0.004mL -0.049mL +0.002mL +0.008mL +0.006mL +0.022mL +0.003mL
Pipetting repeatability ±0.002mL ±0.004mL ±0.002mL ±0.001mL ±0.004mL ±0.002mL ±0.002mL ±0.002mL
Adjustment reproducibility ±0.003mL ±0.002mL ±0.000mL ±0.002mL ±0.006mL ±0.006mL ±0.008mL ±0.002mL
Overall Evaluation (RMS) Systematic error 0.006 0.006 0.034 0.005 0.006 0.005 0.015 0.005
Pipetting repeatability 0.003 0.004 0.004 0.003 0.005 0.004 0.003 0.003
Adjustment reproducibility 0.003 0.008 0.007 0.007 0.011 0.012 0.016 0.002

4.3 Discussion

Overall, the accuracy of domestic pipettes is impressive. We should discard the prejudice that domestic instruments equate to “low quality.” In fact, they can be excellent cost-effective alternatives.

  • Repeatability: Domestic pipettes are nearly on par with imported ones.
  • Adjustment Reproducibility: Slightly inferior to imported models but within acceptable limits.
  • Linearity: Performance is inconsistent; some are close to imported standards, while others show a significant gap.

(1) Systematic Error and Linearity

Since most pipettes only allow for single-point calibration (usually at full scale), errors at other volumes cannot be proportionally corrected if linearity is poor. Most domestic pipettes met the error requirements across the full range, but a few showed large errors at lower volumes. For instance, the BKMAM pipette had a systematic error near $-0.05 \text{ mL}$ at the $0.5 \text{ mL}$ setting, reaching the scale interval value. Ideally, the Maximum Permissible Error (MPE) should not exceed half of the scale interval. We recommend manufacturers introduce two-point calibration to improve linearity.

(2) Adjustment Reproducibility

Domestic pipettes use digital dials, which can lead to the misconception that the displayed number is the absolute volume. However, mechanical threads always have clearances (backlash). The test data shows that domestic digital dials are less precise than the imported pointer-style dials, indicating that the precision of domestic adjustment threads needs improvement.

(3) Piston Rebound and Durability

When the piston is too tight and spring tension is insufficient, the piston may fail to return fully, leading to under-pipetting. Under high-intensity use, metal fatigue in the spring can cause temporary failure. As shown in Figure 4, performing more than 10 continuous aspirations without pause revealed a gradual decrease in actual volume, with the variance exceeding $3\times$ the normal SD. This phenomenon was observed across five domestic brands. Interestingly, the BKMAM (due to smooth piston movement) and Fenyi Sihai (due to high spring force) did not exhibit this issue.

  1. Conclusion
  1. To Manufacturers: Pipettes are small but vital. Domestic brands are cost-effective, but minor quality gaps remain. Manufacturers should focus on “high-quality development” and precision rather than just low-price competition.
  2. General Recommendation: Domestic pipettes offer great value. While minor deficiencies exist, they provide satisfactory results if used and calibrated correctly. Given potential issues at the extreme ends of the range, we suggest using the middle range of the volume scale and ensuring regular calibration.
  3. Closing Thought: The conclusion regarding spring fatigue and decreasing volume is somewhat speculative and requires more data. An alternative theory suggests that heat transfer from the palm during prolonged holding affects the volume. However, this does not explain why only certain brands or specific (full-scale) volumes are affected, especially since the piston is located at the bottom, relatively far from the handgrip.

It’s one of the most essential tools in any lab, but do you know all its types and functions? Dive into our ultimate guide on pipettes-What is a pipette.