Combined Pharmaceutical Stability Test Chamber — Multi-chamber Series

Product Features

  • Combines stability and lighting tests to shorten testing time, save floor space and avoid equipment changes during testing;
  • A special antibacterial stainless-steel chamber with rounded corners is easy to clean and convenient to use;
  • New CFC-free design offers high efficiency and low energy consumption to promote energy savings;
  • An LCD touchscreen controller displays all functions graphically and in text, with curves showing equipment operating status;
  • Continuous operation: two imported hermetic compressors switch automatically to ensure prolonged continuous operation without faults;
  • Continuous operation requires no defrosting, avoiding chamber temperature and humidity fluctuations caused by defrosting during use;
  • An imported humidity sensor avoids frequent wick replacement associated with conventional wet- and dry-bulb sensors;
  • Independent temperature, humidity and lighting control in each chamber shortens testing time and improves efficiency when researching and developing new drug stability;

Tiered Access Management

Programmable Touchscreen Controller (Refer to SDP/GSP Series)

  • Multiple individual accounts can be assigned according to equipment-management requirements, with three controller access levels: administrator, operator and guest. The touchscreen supports Chinese and English input, allowing users to log in under their actual names. Detailed user-operation logs can also be queried for equipment operation, maintenance and audit trails;

Safety Functions

  • Air circulation and heating stop automatically when the door opens to prevent temperature overshoot;
  • Automatic circulating fan speed control prevents sample evaporation caused by excessive airflow;
  • Password-protected screen locking prevents incorrect operation by non-laboratory personnel;
  • An independent temperature-limit alarm system provides audible and visual warnings;
  • Temperature-deviation alarms; compressor overheating, overload and overpressure protection; water-shortage protection;

3Q qualification and calibration services are available.

Technical Specifications

Model LHH-SS-I (two chambers) LHH-SS-II (two chambers) LHH-SG-I (two chambers) LHH-SG-II (two chambers) LHH-SSG (three chambers)
Temperature Control Range 0~65°C 0~65°C (15~50°C with lighting) 0~65°C (15~50°C with lighting)
Temperature Fluctuation ±0.4°C
Temperature Uniformity ±2°C
Humidity Range 25~95%RH
Humidity Deviation ±3%RH
Light Intensity / Error – 0~6000LX adjustable, ≤±500LX 0~6000LX adjustable, ≤±500LX
Refrigeration System Imported compressor unit Imported compressor unit Imported compressor unit
Controller Programmable touchscreen controller
Sensor Temperature: Pt100 platinum resistance sensor; humidity: capacitive humidity sensor
Voltage AC220V 50HZ AC380V 50HZ
Rated Power / Operating Power 3100W / approx. 1200W 3250W / approx. 1200W 5200W / approx. 2000W
Capacity (L) Upper Lower Upper Lower Upper Lower Upper Lower 1 2 3
100L 100L 150L 150L 100L 100L 150L 150L 100L 100L 225L
Chamber Dimensions
W×D×H (mm)
500×450×450 660×520×450 500×450×450 660×520×450 500×450×450 500×450×1000
Overall Dimensions
W×D×H (mm)
685×743×1835 845×743×1970 685×743×1835 845×743×1970 1300×801×1835
Lighting – – – – √ – √ – √ – –
Shelves (Standard) 2 2 2 2 2 2 2 2 2 2 4

Notes: 1. Ambient operating temperature: +5~30°C.
2. For LHH-SS, LHH-SG and LHH-SSG, delivery is 15 days after receipt of the order.
3. LHH-SS consists of chambers 1 and 2; LHH-SG consists of chambers 1 and 2; LHH-SSG consists of chambers 1, 2 and 3. Chamber 1 of LHH-SG and LHH-SSG has lighting.