Description

Application:

BK-5100 series is a high-end product, suitable for research and development.

Features:

*Highly integrated ARM microcontroller SPI, I2C technology

* Accurate automatic wavelength positioning and excellent wavelength accuracy

* Plunger floating design, more stable output

* Intelligent sequence analysis and batch processing

* Chromatography workstation software meets 21 CFR Part 11

Technical Parameters:

ModelBK-LCI5100BK-LCG5100BK-LCQL5100
PumpTypeIsocratic PumpGradient PumpQuarternary Low Pressure Pump
Flow Rate Range0.001-10.000mL/min, in 0.001mL increments
Flow Accuracy≤±0.2% (1.00ml/min, H2O,R.T.,8.5MPa)
Flow PrecisionRSD≤0.06% (1.00ml/min, H2O,R.T.,8.5MPa)
Maximum Pressure63MPa
Pressure accuracy≤±3.0%
Pressure pulsation≤0.1MPa
Composition rang/0-100%0-100%
Composition precision/≤0.2%RSD≤0.3%RSD
UV/VIS

Detector

Wavelength Range190-800nm
Wavelength Repeatability≤0.1nm
Wavelength Accuracy≤±1.0nm
Band Width6nm
Response Time0.0s-5s
Baseline Noise≤±0.25×10-5AU(empty cell, 254nm, 1.0s)
Drift≤0.5×10-4AU/h(empty cell, 254nm, 60min)
Path length10 mm(analytical cell)
Light SourceDeuterium Lamp and Tungsten Lamp
Data-processing Workstation◆ Sequence tables for any set of samples

◆ Internal and external standard calculation methods

◆ Automatically displays, prints, exports and starts other programs after the completion of a measurement.

◆ Comply with the requirements of the 21 CFR Part 11 directive of the FDA

5100 System Kits1 Set
5100 System ManagerIntegrated A/D convertersIntegrated A/D converters

Integrated vacuum degasser, 4 channel

Power SupplyAC110/220V±10%, 50/60Hz
Optional AccessorySolvent bottle,Column Oven, Auto-sampler,DAD512 Detector, Refractive Index Detector, Evaporative Light Scattering Detector, Fluorescence Detector, PC
Package Size(W*D*H)700*570*310mm; 700*570*350mm; 710*583*500mm;

730*450*440mm; 600*440*380mm; 770*450*700mm;

Gross Weight156kg