I. Core Selection Factors
1. Define Testing Requirements
- Wavelength range: Standard 190–1100nm (budget option: 325–1000nm)
- Sample type: Liquid (standard cuvette) / Solid (integrating sphere) / Micro-volume (micro cell)
- Accuracy: 1–2nm bandwidth for qualitative analysis; <0.5nm for quantitative
2. Key Configuration Comparison
Feature | Basic/Teaching | Research-Grade | Use Case |
---|---|---|---|
Light Source | Deuterium + Tungsten | Xenon (longer lifespan) | Continuous operation |
Detector | Silicon photodiode | PMT (10× sensitivity) | Trace analysis |
Optical System | Grating (1200 lines/mm) | Double monochromator | Stray light reduction |
Data Interface | USB/local storage | LIMS integration | GMP labs |
3. Special Functions
- Kinetics: Time-scan (min. interval: 0.1s)
- Multi-component: Multi-wavelength/full-spectrum scan
- Ultra-trace: Fiber optic probe (≤50μL volume)
II. International vs. HINOTEK Comparison
Parameter | International Brand | Chinese Spectrophotometer (HINOTEK) |
---|---|---|
Price Range | US$10000.00~US$80000.00 | <US$300.00 (721G Visible Spectrophotometer) <US$800.00 (752N UV-Visible Spectrophotometer) <US$2000.00 (L7 Double Beam UV-Visible Spectrophotometer) |
Wavelength Accuracy | ±0.1nm | ±0.2–0.5nm |
Stray Light | <0.01%T (220nm) | <0.1%T |
Scan Speed | 5000nm/min | 2000nm/min |
Recommended Use | Pharmacopeia/research | Teaching/routine QC |
III. Sample Preparation & Measurement Tips
1. Liquid Samples
- Solvent: Ensure transparency at target wavelength (e.g., quartz cuvette for 260nm)
- Concentration: Optimal absorbance 0.2–0.8
- Debubbling: Settle or low-speed centrifugation
2. Cuvette Guidelines (View HINOTEK cell)
Material | Wavelength Range | Notes |
---|---|---|
Quartz | 190–2500nm | Avoid hydrofluoric acid |
Glass | 340–2500nm | Resists strong alkalis |
Plastic | 380–780nm | Single-use only |
3. Special Samples
- Turbid samples: Centrifuge or filter (0.22μm)
- High absorbance: Dilute or use short-path cuvette (e.g., 1mm)
IV. Instrument Optimization
1. Parameter Setup
- Scan mode: Fixed wavelength (quantitative) / Full spectrum
- Bandwidth: 2nm (routine); <1nm (sharp peaks)
2. Baseline Correction
- Blank solvent calibration (mandatory for solvent changes)
- Dual-beam compensation (reduces solvent noise)
3. Error Prevention
- Align cuvette markings (frosted side toward light)
- Warm up for 30 minutes before use
- Dilute high-concentration samples (A >1.0 causes nonlinearity)
V. Data Analysis & Validation
1. Quantitative Workflow
- Calibration curve: 5–7 points, R² >0.999, QC verification
- Multi-component: Ensure Δλ >30nm to avoid overlap
2. Spectral Peaks
- Proteins: 280nm (tryptophan/tyrosine)
- DNA/RNA: 260nm
- Aromatics: Multi-peaks (200–300nm)
3. Method Validation
- LOD/LOQ: S/N = 3/10
- Precision: RSD <1% (6 replicates)
VI. Maintenance & Troubleshooting
1. Routine Care
Frequency | Task | Standard |
---|---|---|
Daily | Clean cuvette windows | No fingerprints/scratches |
Weekly | Light source energy check | >80% initial output |
Quarterly | Wavelength calibration | Holmium oxide standard |
2. Common Issues
Symptom | Cause | Solution |
---|---|---|
Noisy baseline | Lamp aging/power fluctuation | Replace lamp/use stabilizer |
Drifting readings | Cuvette contamination | Ultrasonic clean (1% HNO₃) |
Wavelength shift | Grating misalignment | Calibrate with Hg lamp |
3. Component Lifespan
- Deuterium lamp: 1K–2K hours (avoid frequent on/off)
- Tungsten lamp: 5K hours
- Cuvettes: Quartz (2 years); glass (1 year)
VII. Safety & Compliance
- UV protection: Close sample compartment; wear UV-blocking goggles
- Chemical safety: Use sealed cells for corrosive solvents; segregate waste
- Electrical: Wait 5 minutes after power loss before restarting
Typical Applications
- Protein quantification: Bradford assay (595nm) – avoid detergent interference
- Water COD testing: 620nm (dichromate method) or UV 254nm (correlation required)
- Nanoparticle analysis: Full scan + Mie theory sizing
Pro Tip: Proper blank correction and cuvette handling eliminate 80% of errors. Establish SOPs, especially for pharmacopeia compliance (wavelength accuracy, absorbance repeatability).