UV-1700S UV Visible Spectrophotometer

Product description, technical parameters and configuration

Introduction

UV-1700S successful implementation of the stringent requirements of high accuracy and reliability measurement to meet the requirements of various applications that can be used in biological research, bio-industry, pharmaceutical analysis, pharmaceutical, teaching and research, environmental protection, food hygiene, clinical examination, health and epidemic prevention and other fields.Huge memory space to store multiple sets of data and curves

A wide wavelength range, the wavelength range to meet the various requirements of the field.

UV-1700S uses 4nm, 2nm, 1nm, 0.5nm four spectral bandwidth to meet the strict requirements of national pharmacopoeas and different users

The design of LSI greatly improves the scalability and reliability of the system

Improved optimization of the optical design, import source and receiver system created a high performance and high reliability.

Rich measurement methods, with a wavelength scan, time scan, multi-wavelength determination, multi-order derivative determination (optional), dual-wavelength, three-wavelength (optional) DNA protein measurements (optional), and other measurement methods to meet different measurement requirements, and can be displayed directly on a 6-inch screen.

Based on user requirements optional hole rack, manual four with stand, manual Eighth rack, automatic Eighth frame, glass stand, test tube rack, 1cm colorimetric frame, 5cm colorimetric frame

Support external U disk storage and transmission of experimental data

Can be powered down to save the measurement parameters and data, user-friendly.

Can achieve more accurate and flexible measurement via PC control, which can meet the needs of different users.

Optical Specifications

Optical path structure:Double beam

Wavelength Range:190-1100nm

Band Width:0.5/1.0/2.0/4.0nm adjustable

Wavelength Accuracy:±0.3nm

Wavelength Repeatability:0.2nm

Photometric Accuracy:±0.3%T

Photometric Repeatability:0.2%T

Stray Light:≤0.03%T

Stability:0.001A/h@500nm

Photometric Mode:T,A,C,E

Wavelength Setting:Automatic

Photometric Display Range:-4~4A

Display Mode:6 inches high brightness blue LCD

Detector:Import Silicon Photodiode

Light Source:Import Deuterium Lamp & Tungsten Halogen Lamp

Power Requirement:AC 220V/50Hz 110V/60Hz

Power:120W

Dimensions(W*D*H):560×450×230mm

Weight:28Kg