It is used to make determination of the closed cup flash point of the petroleum products.
Features
BK-FP261
* The heating power is adjustable and shown visually by voltmeter.
* The structure is designed small and exquisite with stainless-steel table board
BK-FP261A
* Simulation tracking displays function curves of heating up and test time.
* Automatically calibrate the effect of atmospheric pressure to test the corrected value.
* Differential coefficient detection.Automatically correct the system deviation.
* Automatically open the lid,detect the flash point and print test data.
* The test arm rises up and lowers down automatically.
* Igniting with electric igniting gun,forced air cooling.
odel | BK-FP261 | BK-FP261A | |
Ambient Temperature | ≤ 35℃ | ≤ 40℃ | |
Relative Humidity | ≤85% | ≤80% | |
Consumption | 650W | 500W | |
Flash Point Determination |
Mercury-in-glass thermometer.Specifications are as below:
(1)Scale -5℃~110℃, division 0.5℃ (2)Scale 20℃~150℃, division 1℃ (3)Scale 90℃~370℃, division 2 |
Range: Ambient to 300℃
Repeatability: ≤3℃ Reproducibility: ≤4℃ Accuracy: 0.1℃ |
|
Oil Cup |
– Inner diameter: 50.7mm~50.8mm
-Depth: 55.7mm~56.0mm – The scribed line depth of capacity of testing oil: 33.9mm~34.3mm -Capacity of testing oil: about 70ml |
||
Stirring Rate | Procedure A: (90~120)RPM, Procedure B: (250±10)RPM | ||
Igniting Device
|
(1) Igniting source: gas(or other civilian fuels, the same below)
(2) Electric ignition. Gas flame diameter: 3.2mm~4.8mm |
||
Heating Device | The heating power is continuously adjustable from 0 W to 600W | ||
Heating Rate | (1~12)℃/min; Controllable and adjustable | Procedure A: (5~6)℃/min,
Procedure B: (1~1.6)℃/min |
|
Power Supply | AC220V±10%, 50Hz | ||
Dimension(W*D*H) | 340*320*450mm | 520*360*310mm | |
Packing Size (W*D*H) | 470*400*410mm | 600*510*600mm | |
Gross Weight | 12.5kg | 25kg |