ANBH torque wrench tester

PRODUCT DESCRIPTION

Main uses and scope of application

ANBH torque wrench tester is a special equipment for testing torque wrenches and torque drivers. Mainly used for testing or calibrating torque wrenches, preset torque wrenches, pointer torque wrenches. Widely used in electrical manufacturing, mechanical manufacturing, automotive light industry, professional scientific research and testing industries. The torque value is displayed through a digital meter, which is accurate and intuitive.

Product features

This instrument is a digital torque wrench tester specially designed for calibrating or adjusting torque wrenches.

(1) Multiple units can be selected

Units of measurement include Newton units (N.m), metric units (kgf.cm) and US units (lbf.in).

(2) Two measurement modes

The real-time and peak modes of this instrument can be switched at will.

(3) Upper and lower limit settings

The upper and lower limits of this instrument can be set. When the set value is reached, the buzzer will alarm.

(4) Data saving function

This instrument can save 100 sets of measurement data.

Variety specifications

Second, the overall product structure

specification ANBH-20 ANBH-50 ANBH-100 ANBH-200 ANBH-500
Maximum 20N.m 50N.m 100N.m 200N.m 500N.m
Minimum resolution 0.001 0.001 0.001 0.01 0.01
Accuracy ± 1%
Unit switch N.m, Kgf.cm, Lbf.in
power supply Input: AC 220v Output: DC 12V
Dimensions mm 299 * 299 * 242 772 * 400 * 276 1142 * 553 * 360
Operating temperature 5 ℃ ~ 35 ℃
Transport temperature -10 ℃ ~ 60 ℃
Relative humidity 15% ~ 80% RH
working environment No vibrating sources and corrosive media around
Weight (Kg) 12 27 43

Key introduction

  •  Up key: Press this key on the measurement interface to rotate the measured data by 180 °.
  • Down key: In the setting interface, it has the function of down selection and displacement.
  • Switch key: power on and off function.
  • Clear key: The measurement interface has data clear function, and the setting interface has return function.
  • View key: In the measurement interface, there is a function to view saved data.
  • Mode selection: In the measurement interface, press this key to switch between peak mode and real-time measurement mode. Long press this key for 6 seconds to enter the setting interface. This key has a save function in the setting interface.

 