Features:
* Auto 100%T/0A and 0%T.
* Test results are displayed by a Large digital LCD screen, digitally from 0 to 100% Transmittance or 0 to 1.99 Absorbance over a continuous wavelength range of 360 to 1000nm.
* Sample compartment contains both round test tube and square cuvette (require an adapter).
* Silicon photodiode detector covers the entire wavelength, no need to change detector.
* 10nm spectral bandwidth provides the sensitivity and performance for almost all applications.
* Built-in stray light filters. Able to test micro samples.
Specifications: |
Model |
SP1102 |
Optical system |
Littrow Mounting, single beam, 1200l/mm grating optical system. |
Wavelength range: |
360-1000nm |
Spectral bandwidth: |
10nm |
Wavelength accuracy: |
± 2.5nm |
Wavelength readability: |
± 1nm |
Stray radiant energy: |
Less than 2% at 340nm |
Photometric range: |
0-100%T, 0-1.99A |
Photometric accuracy: |
± 3%T |
Light Source: |
6V/10W Tungsten Halogen Lamp |
Sample Container: |
Φ12mm test tube or 10*10*45mm rectangular cuvette ( require an adapter) |
Detector: |
High sensitivity silicon photodiode |
|
| Standard Accessories: Instruction manual, Dust cover, 10*10*45 rectangular cell |
|
: |
|
|