Microscope Oven & Incubator Pipette Spectrophotometer Photometer & Colorimeter Cell (Cuvette) Hand Held Refractometer Viscometer Polarimeter Abbe Refractometer Centrifuge Melting point machine Balance Distillation Apparatus Autoclave & Sterilizer Mixer, Stirrer & Shaker PH & Conductivity Meter Vacuum Blood Tube Hemoglobinometer, Hemocytometer & Colonometer Water Bath,Oil &Dry Bath Telescope Titrator, Ion Meter & D.O.ANALYZER    Automatic Potential Titrator    Ion Meter    D.O.Analyzer    Multi-parameter Analyzer Muffle Furnace Sample Concentration Flame Photometer Biological Safety Cabinet Laminar Flow Cabinet

Catalog and Manual

Catalog

Download E-catalog

Require Original Catalog

Require E-manual


image
Titrator, Ion Meter & D.O.ANALYZER arrow Automatic Potential Titrator arrow ZDJ-5 Automatic Potential Titrator

ZDJ-5 Automatic Potential Titrator
 
ZDJ-5
Characteristics:
1.Modular design consisting of three sections: capacity titration unit (can be replaced by Coulomb titration unit), control unit ( can be replaced by PC titration software) and measurement unit (including measurement of potential, conductivity and constant) to realize potential titration, coulomb titration, conductivity titration and constant titration.
2.Touch screen control, easy to operate.
3.Titration software control to realize all kinds of operation.
4.Electricity cutting off protection.
5.Special titration modes of pre-titration, pre-setting terminal titration, empty terminal titration or manual titration to expand the range of application.
6.Titration system made of perchloric acid resistant materials to make nonaqueous titration possible.
 
 
Specifications:
1. Measuring range:      pH:  0.00~14.00pH
                                      mV: 0~±1999.0mV
                                       temp.: -5~105°C
2. Accuracy of the electric unit  pH: ±0.01pH±1 bit
                                                        mV: ±0.03%FS
                                                        temp.: ±0.3°C
3. Input current: Not greater than 1´10-12 A
4. Input impedance: Not less than 3´1012 W
5. Stability: ±0.3mV±1 bit/3h
6.Control titration sensitivity: ± 2mV
7. Titration tube capacity error: 10ml tube: ±0.025ml
                                                        20ml tube: ±0.035ml
8. Titration rate: 55±10s (to the full scale of the tube)
9. Capacity analysis reproducibility: 0.2%.
: