laboratory
Tuesday, 07 February 2012
top left pic
top pic
equipment
 
laboratory
Home arrow Products arrow Titrator, Ion Meter & D.O.ANALYZER arrow Multi-parameter Analyzer arrow DZB-718 Portable Multi-Parameter Analyzer
logo right
Products
Microscope Spectrophotometer Oven & Incubator PH & Conductivity Meter Titrator, Ion Meter & D.O.ANALYZER    Automatic Potential Titrator    Ion Meter    D.O.Analyzer    Multi-parameter Analyzer Pipette Centrifuge Photometer, Colorimeter & others Viscometer Polarimeter Abbe Refractometer Hand Held Refractometer Melting point machine Balance Distillation Apparatus Autoclave & Sterilizer Mixer, Stirrer & Shaker Vacuum Blood Tube Hemoglobinometer, Hemocytometer & Colonometer Water Bath,Oil &Dry Bath Muffle Furnace Sample Concentration Flame Photometer Biological Safety Cabinet Laminar Flow Cabinet Cell (Cuvette) Gas Chromatograph Diaphragm Vacuum Pump

Catalog and Manual

Catalog

Download E-catalog

Require Original Catalog

Require E-manual


image
Titrator, Ion Meter & D.O.ANALYZER arrow Multi-parameter Analyzer arrow DZB-718 Portable Multi-Parameter Analyzer

DZB-718 Portable Multi-Parameter Analyzer

Main Character:
* Display several parameter's measuring value at the same interface. Support to measure pX/pH, mV, conductivity, TDS, salinity, resistivity, DO, saturation, DO currency and temp. value simultaneously.
* Provide several common ion modes randomly, such as H + , Ag + , Na + , K + , NH4 + , Cl - , F - , NO3 - , BF4 - , CN - , Cu2 + , Pb2 + and Ca2 + . As long as user has relevant ion electrode, user could set up his own ion mode.
* Can view, print and delete measuring data.
* With USB interface, intelligent power management. If match with special communication software, it can be connected with PC.
* Three measuring modes: continuous measuring mode, timed measuring mode and balance measuring mode.
* Support GLP regulation, can print and store 200 sets for each of pH, conductivity, TDS, salinity and DO. Allow to store 6 kinds of ion measuring value, including relevant pX value and ion concentration value 100 sets for each.
* Small in size, low power dissipation, IP65 protection class for easy moving and field measurement.
 
Main Specification:
* Measuring Range:
a). mV: (-1999.99 ~ 0)mV, (0 ~ 1999.99)mV;
b). pH/pX: (-2.000 ~ 19.999)pH/pX;
c).Concentration: various concentration values in correspondence to potential measuring range and indicating electrode.
d). Conductivity: 0.000μS/cm ~ 199.9mS/cm
e). Resistivity: 5.00Ω.cm ~ 20MΩ.cm
f). TDS: 0.000mg/L ~ 19.99g/L
g). Salinity: (0.0 ~ 8.00)%
h). Dissolved Oxygen Concentration: (0.00 ~ 19.99)mg/L
i). Dissolved Oxygen Saturation: (0.0 ~ 199.9)%
j). Temperature: (-5.0 ~ 135.0) °C
 
* Resolution :
a). pH/pX: 0.01/0.001pH
b). mV: 0.1/0.01mV
c). Dissolved Oxygen: 0.01mg/L
d). Saturation: 0.1%
e). Temperature: 0.1°C
 
* Electronic Unit Accuracy:
a). pH/pX: ±0.002pX ± 1 bit
b). mV: ±0.03% (FS)
c). Ion concentration: ±0.5% ± 1 bit
d). Conductivity: ±0.5% (FS) ± 1 bit
e). TDS: ±0.5% (FS) ± 1 bit
f). Salinity: ±0.1% ± 1 bit
g). Dissolved Oxygen: ±0.10mg/L
h). Dissolved Oxygen saturation: ±1.0%
i). Temperature: ±0.3°C ± 1 bit
 
* Input Impedance: >3*10 12Ω
* Power supply: 4 pieces AA alkaline cell.
* Outside Dimension (mm):210*100*45
* Weight: about 0.5kg
:
pic
email to us