| Characteristic: |
| 1. | Dot-matrix LCD. |
| 2. | Measure mV value,pH value, PX value |
| 3. | Suitable for measuring solution concentration by way of concentration immediate reading,known adding method,sample adding and GRAN method. |
| 4. | Measurement results can be stored,deleted, consulted and printed.It can store at most 50sets of measurement data.The user has two ways to select immediate printing or storage printing. |
| 5. | With four slope calibration methods,including one point calibration,two-point calibration,multi-point calibration and multi adding calibration. |
| 6. | Electricity cutting off protection.The stored data and parameters will be kept even if unusual electricity cutting off happens; |
| 7. | With RS-232 interface,TP-16 printer can be connected. |
| | |
| |
| Specification: |
| 1. | Measurement range: |
| | PX±1 or pX±2: 0.000-13.999 pX±1 or pX±2 |
| | mV:-1800.0-1800.0mV |
| | Temperature: -5-105.0 °C |
| 2. | Resolution: |
| | PX±1 or pX±2: ±0.001pX |
| | mV:0.1mV |
| | Temperature: 0.1 °C |
| 3. | Accuracy of the electric unit: |
| | PX±1 or pX±2: ±0.005pX ± 1 bit |
| | MV: ±0.03%(FS) ±1 bit |
| | Temp: ±0.3 °C ± 1 bit |
| 4. | Stability of the electric unit: ±0.005pX ± 1 bit/3h |
| 5. | Repeatability: 0.01pX |
| 6. | Temperature compensantion:0-60.0°C |