Online Service
laboratory
Thursday, 09 September 2010
top left pic
top pic
equipment
 
laboratory
logo right
Products
Microscope Oven & Incubator Pipette Spectrophotometer Photometer & Colorimeter Cell (Cuvette) Hand Held Refractometer Viscometer Polarimeter Abbe Refractometer Centrifuge Melting point machine Balance Distillation Apparatus Autoclave & Sterilizer Mixer, Stirrer & Shaker PH & Conductivity Meter Vacuum Blood Tube Hemoglobinometer, Hemocytometer & Colonometer Water Bath,Oil &Dry Bath Telescope Titrator, Ion Meter & D.O.ANALYZER    Automatic Potential Titrator    Ion Meter    D.O.Analyzer    Multi-parameter Analyzer Muffle Furnace Sample Concentration Flame Photometer Biological Safety Cabinet Laminar Flow Cabinet

Catalog and Manual

Catalog

Download E-catalog

Require Original Catalog

Require E-manual


image
Titrator, Ion Meter & D.O.ANALYZER arrow D.O.Analyzer arrow JPSJ-605 D.O. Analyzer

JPSJ-605 D.O. Analyzer
 
JPSJ-605
Characteristics:
1. Microprocessor technology. Dot-matrix LCD with blue-light.
2. With the functions of automatic temperature compensation and calibration of zero oxygen, full scale, atmosphere pressure and salinity
3. With the function of double displays of dissolved oxygen concentration value and temperature, dissolved oxygen saturation value and temperature
4. With Measuring results can be stored, deleted, cousulted and printed. At most 50 sets of data can be stored. the user can two ways to select immediate printing or storage printing.
5. The meter has the function of protection from electricity cutting off;
6. With RS-232 interface which is connected and with Model TP-16 printer to print measuring results or connected with computer to communicate;
7. Using 1.1 communication soft can connect with compute
 
 
Specifications:
1. Measurement range:    Dissolved oxygen: 0.00-20.00 mg/L
                                           Dissolved oxygen saturation: 0.00-200.0 %
                                          Temperature: -5.0-105.0°C
2.The range of automatic temperature compensation: 0-40°C
3.Accuracy:
    Dissolved oxygen: ±0.3 mg/L± 1 bit (under the basic temperature)
                                      ±0.5 ml/L± 1 bit (differ: ±15°C from the basic temperature)
   Dissolved oxygen saturation: ±5%± 1 bit (under the basic temperature)
                                                         ±10%± 1 bit (differ: ±15°C from the basic temperature)
   Temperature: ±0.5°C ± 1 bit (0°C £T £ 60°C)
                             ±1.0°C ± 1 bit (on the other temperature range)
4. Remaining current: ≤0.15mg/L
5. Response time: 45s (90% response on 20°C)
6. Salinity calibrating range: (0.0~40.0)g/L
7. Repeatability of the meter: ≤0.2mg/L.
8.Stability of the meter: not over ±0.2mg/L within one hour.
:
pic
email to us