Microscope Spectrophotometer Oven & Incubator PH & Conductivity Meter Titrator, Ion Meter & D.O.ANALYZER    Automatic Potential Titrator    Ion Meter    D.O.Analyzer    Multi-parameter Analyzer Pipette Centrifuge Photometer, Colorimeter & others Viscometer Polarimeter Abbe Refractometer Hand Held Refractometer Melting point machine Balance Distillation Apparatus Autoclave & Sterilizer Mixer, Stirrer & Shaker Vacuum Blood Tube Hemoglobinometer, Hemocytometer & Colonometer Water Bath,Oil &Dry Bath Muffle Furnace Sample Concentration Flame Photometer Biological Safety Cabinet Laminar Flow Cabinet Cell (Cuvette) Gas Chromatograph Diaphragm Vacuum Pump

Catalog and Manual

Catalog

Download E-catalog

Require Original Catalog

Require E-manual


image
Titrator, Ion Meter & D.O.ANALYZER arrow D.O.Analyzer arrow JPBJ-608 D. O. Analyzer

JPBJ-608 D. O. Analyzer
 
JPBJ-608
Characteristics:
1. With the function of double displays of dissolved oxygen concentration value and temperature, dissolved oxygen saturation value and temperature
2.Measuring D.O. concentration, saturation, electrode current and temperature; With the functions of automatic temperature compensation and calibration of zero oxygen, full scale, atmosphere pressure and salinity.
3.The meter has the function of protection from electricity cutting off. With Measuring results can be stored, deleted, viewed and printed. At most 250 sets of data can be stored.
4. With wide screen LCD. It is easy to operate with the prompt facility. With PC panel of new material, thus the meter has good reliability.
5. With low power consumption design, with the functions of lack of voltage display, switching off automatically.
6.With RS-232 interface which is connected and with Model TP-16 printer to print measuring results or connected with computer to communicate.
7.Grade of the case protection is IP65, resistant to water and dirt, suitable for field operation
8.The electrode system is composed of polarographic oxygen electrode and temperature electrode.
 
 
Specifications:
1.Measuring range: Dissolved oxygen concentration: (0.00~19.99)mg/L
                                Dissolved oxygen saturation:   (0.0~199.9)%;
                                                               Temperature: (0~40.0)℃.
2.Automatic temperature compensation range:      (0.0~40.0) ℃.
3.Accuracy of the meter: D.O. concentration: ±0.3mg/L±1bit(under basic temperature);
                                                                       ±0.5mg/L (differ: ±15°C from the basic temperature)
                                             D.O. saturation :  ±5.0% (under basic temperature);
                                                                            ±10.0% differ: ±15°C from the basic temperature
                                             Temperature:  ±0.5℃±1bit.
4.Repeatability of the meter: ≤0.2mg/L.
5.Stability of the meter: not over ±0.2mg/L within one hour.
6.Remaining current: ≤0.1mg/L.
7.Response time ≤20s (90% response at 20.0℃).
8.Salinity calibrating range: (0.0~40.0)g/L.
9.IP65 protection grading
: