DDS-307 Conductivity Meter | | | | | Characteristics: | | 1.31/2LED display | | 2.With manual temperature compensation. | | 3.There are four types of electrodes for conduction cell constant: 0.01,0.1,1.0 and 10.0cm-1 ,user can select by your need. | | 4.Automatic switching measuring frequency, measuring ultrapure water with using K=0.01cm-1 titanium alloy electrode and sealing measuring cell | | 5.With 0~10mV output signal | | | | | | | | Specifications: | | 1.Measurement range: 0-1×105 mS/cm | | There is the best correspondence point between conductivity measuring range and conduction cell constant |
| conduction cell constant(cm-1 ) | 0.01 | 0.1 | 1 | 10 | | Measuring range (uS/cm) | 0~2 | 0.2~20 | 2~1*104 | 1*104~1*105 |
| 2.Electrode constant, cm-1: 0.01. 0.1, 1.0.10.0 | | 3.Accuracy of the electric unit: ±0.5%(FS) ±1 bit | | 4. Stability of the electronic unit: ±0.33%(FS) ±1bit/3h | | 5.Accuracy ±1.0%(FS) ±1 bit | | 6.The range of temperature compensation: 15-35°C, basic temperature: 25° |
| | | Standard Configuration: | | 1. Main instrument 1; | | 2. Power cable 1; | | 3. Platinum Electrode 1; | | 4. Multi-function Electrode Support Stand 1; | | Optional accessories: | | 1. DJS-1 bright Electrode | | 2. DJS-10 Conductivity Electrode | | 3. DJS-0.1 Platinum Electrode | | 4. DJS-0.01 Titanium Alloy Electrode | | Overall dimension & weight: | | Shipping package dimension: 332-265-275mm; | | Gross weight: 3kgs; | | Net Weight: 2kgs. |
|
: |
|
|