DDS-11A Conductivity Meter |
|
| Functions: |
| * With 3 1/2 LED, display is clear and precision is high. |
| * Electrode constant compensation. |
| * Manual temperature compensation. |
| * Output signal (0-10)mV. |
| |
| 1.Measuring range: |
| The measuring range is (0~2×105) uS/cm. The instrument has five ranges and can be changed through band switch by manual (by electrode with constant). |
Range | Measuring range | Resolution | 2 uS/cm | 0.001~2 uS/cm | 0.001 uS/cm | 20 uS/cm | 0.01~20 uS/cm | 0.01uS/cm | 200 uS/cm | 0.1~200 uS/cm | 0.1uS/cm | 2 mS/cm | 0.001~2.mS/cm | 0.001mS/cm | 20 mS/cm | 0.01~20 mS/cm | 0.01mS/cm |
|
| NOTE: To measure high conductivity, a larger constant electrode is commonly used. If the conductivity is over 10000 uS/cm, the electrode with constant of 10 cm―1 is used and the measuring range is extended to 2×105 uS/cm. |
| |
| 2. Accuracy of the electronic unit : ±1.0%(FS) ±1bit |
| 3.Accuracy of the meter: ±1.5%(FS) ±1bit |
| 4. Range of Temp. compensation: 5-35℃ |
| 5.Normal operation conditions: |
| (1) Ambient temperature : (5 – 35)℃ |
| (2) Relative humidity : not more then 85% |
| (3) Power source: AC220V±22V, 50HZ±1HZ |
| (4) Without any notable vibration |
| (5)No strong electromagnetic interference to surrounding except the geomagnetic field. |